Key Points:
•
Reduce VBW can make the spectral line be smooth, and the small signal which is submerged
in noise can stand out. But the scan time will be longer. (When scan time is automatic, it is
influenced by RBW and VBW together).
•
When VBW is automatic, it changes with RBW. But it is not influenced by RBW when
manually.
•
You can use numeric keys, knobs or direction keys to modify this parameter. The detail
method refers to the specification in “Parameter Setting”.
VBW Reference
Parameter
Instruction
Default
1 MHz
Range
1 Hz to 1 MHz
Unit
GHz, MHz, kHz, Hz
Knob Step
in 1, 3, 10 sequence
Direction Key Step
in 1, 3, 10 sequence
V/R Ratio
Set the ratio of VBW and RBW.
Key Points:
•
Choose the ratio according to different signals:
•
When measure sine signals, normally choose 1~3 (to get shorter scan time). When measure
pulse signals, choose 10 (to reduce the influence to the amplitude of nstantaneous signal).
When measure noise signals, normally choose 0.1 (to get the average value of noise).
•
You can use numeric keys, knobs or direction keys to modify this parameter. The detail
method refers to the specification in “Parameter Setting”.
V/R Ratio Reference
Parameter
Instruction
Default
1 MHz
Range
0.000001 ~ 30000
Unit
N/A
Knob Step
in 1, 3, 10 sequence
Direction Key Step
in 1, 3, 10 sequence
3.2.2 Sweep
Set parameters about the Sweep and Trigger functions, including Time, Auto SWT, Mode,
Numbers, Trigger type, etc.
Mode
Set the sweep mode to “Single” or “Continue”, the default is “Continue”. The corresponding icon of
the sweep mode will be displayed in the status bar at the left side of the screen.
1. Single
Set the sweep mode to “Single”. The number
10
on the parameter icon denotes the current sweep
Содержание DSA2016TG
Страница 1: ...USER S MANUAL Handheld Spectrum Analyzer DSA2016TG V1 0 0 ...
Страница 7: ...Chapter1 Quick Start u Instrument Introduction u General Check u Preparation ...
Страница 15: ...2 3 2 Left Panel Left Panel 5 Headphone Connects to a headphone 6 SD card slot 5 6 ...
Страница 19: ...Note in the storage function the direction keys can also be used to select the current path or file ...
Страница 66: ...input 220VAC 15 AC frequency range 40Hz to 60Hz Power consumption Maximum 32W ...
Страница 67: ...Chapter7 Appendix u Appendix A u Appendix B ...