KWE-190107-Rev008-UM-en 

2.10

 

Unit operation 

When scale is on, press [TARE] for more than 3 seconds to change the weight unit between g, 
kg and lbs. Requirement is a stable weight. 
Selected unit will be stored, when scale is switched off.

 

2.11

 

Check weighing – Minimum and maximum values with acoustical alarm 

For setting minimum and maximum values, scale has to be empty. 
Now press both [SET] and [ON] together for 3 seconds.  
Display will show “-no-“ (in delivery condition). Press [TARE], it will circularly display “-no-“,  
“-in-”, “-out-“ and again “-no-“:  
„-no” deactivates the alarm function. 
„-in-” activates the alarm within the range.  
„-out-“ activates the alarm out of the range. 
When display shows “-in-“  or “-out-“, press [ON] to confirm the choice. 

When “-in-“ or “-out-“ is confirmed now the minimum value can be chosen. The display shows 
an arrow at “Under” and the first digit is lighting. The light shows the value set last time and the 
digit which can be adjusted is blinking. Now press [SET] to change the digit, press [TARE] to 
change the digit value, and press [ON] to confirm.  

Then the setting of minimum value is done and the arrow at “Over” is lighting. Use the same 
way to finish the setting of the maximum value. 

When finished setting, it returns to weighing mode and the weighing alarm function is on.  

If the weight is upper the maximum value, the “Over” indicator is lighting, while arrow to “Under” 
is lighting, when weight is under minimum value. 

If the weight is between minimum and maximum values the arrow next to “Accept” is lighting. 

If activated (setting “-in-“ or “-out-“), control weighing is supported by an acoustic signal, which 
sounds, if the weight is in between or out of the minimum and maximum values (as set in the 
parameter; see Cap. 3.8). 

For zeroing minimum and maximum values, set both values back to 000000 or deactivate 
function by choosing “–no-”. 

The settings will be stored in memory when scale is switched off. 

If changing the weighing unit, this function will be closed and the current 
setting value will be auto-cleared. 
 
If changing the graduation, this function will be closed and the current setting 
value will be cleared. 

Содержание KWE Series

Страница 1: ...ss Steel Scales Compact Scale in Stainless Steel Housing Series KWE KWE 190107 Rev008 UM en ADE Germany GmbH Co KG Hammer Steindamm 27 29 22089 Hamburg GERMANY 49 0 40 432776 0 Fax 49 0 40 432776 10 E mail info ade hamburg de ...

Страница 2: ...11 Check weighing Minimum and maximum values with acoustical alarm 5 2 12 Counting Function 6 3 Environment Parameter Setting 6 3 1 Adjustment display precision 7 3 2 Weighing unit kg g lbs 7 3 3 Zero tracking 7 3 4 Stability Filter 7 3 5 Power Saving Mode Standby only for models with rechargeable batteries 8 3 6 Display Update Speed 8 3 7 Creep Tracking Range 8 3 8 Printer setting 9 4 Display Sym...

Страница 3: ...o that there can be stored minimum and maximum values combined with an acoustical signal in the scale e g for weighing always the same weights The stainless steel housing allows using the scale in humid working environments The scale is a precision instrument and must be treated accordingly Therefore kindly read and follow the Operating Manual in all respects 2 1 Check carton contents Compact Scal...

Страница 4: ...d on the adapter 2 4 Power immediate operation Working with mains all three LCD High Mid Low are on For models with rechargeable batteries All three segments lighting Power voltage of rechargeable batteries is more than 6 3V Two segments lighting Power voltage of rechargeable batteries is between 6V und 6 3V One segment lighting Power voltage of rechargeable batteries is under 6V 2 5 Power Charge ...

Страница 5: ...age of the rechargeable batteries is less than 5 6V the scales turn off automatically 2 7 Zero Operation When scale is on ON is the key to zero It is only available within 4 of the maximum capacity otherwise it will immediately display HHHHHH 2 8 Displaying the voltage When scale is on it will display the voltage when pressing ON key for 3 seconds Now if press ON again the scale will return to wei...

Страница 6: ...ET to change the digit press TARE to change the digit value and press ON to confirm Then the setting of minimum value is done and the arrow at Over is lighting Use the same way to finish the setting of the maximum value When finished setting it returns to weighing mode and the weighing alarm function is on If the weight is upper the maximum value the Over indicator is lighting while arrow to Under...

Страница 7: ...n memory when scale is switched off For returning to weighing mode press both SET and ON together for 3 seconds and choose pcs_n by pressing TARE Confirm your choice with ON The setting will be stored and scale return to weighing mode When scale is switched on next time it starts in weighing mode Piece counting function doesn t work if unit lbs is selected 3 Environment Parameter Setting The user ...

Страница 8: ...isplay In this menu the zero tracking can be adjusted if there are weighing performances with weight increasing which are very slow There can be chosen between 11 adjustments 0 5 in half steps by pressing TARE again and again Default is Zer_2 Remark If you have selected a higher graduation PrE 3 1 Adjustment display precision select a high value for stability filter 3 4 Stability Filter Please sta...

Страница 9: ...lay is showing 0 backlight automatically switches off By putting something on the weighing platform again backlight of display will switch on again If nothing will be put on the weighing platform within ten minutes the scale automatically switches off 3 6 Display Update Speed Please start as described in the beginning of this chapter to enter this menu and SPd is shown in the display By pressing T...

Страница 10: ...hargeable battery has been charged fully _Inlt_ Faulty weighing 5 Display Symbols Fault Cause Solution Faulty weighing result Parameter of load cells have changed Re calibration Display shows HHHHHHHH Over zero range Re calibration Load cell or PCB damaged Display shows LLLLLLLL Load cell or PCB damaged Rear display doesn t show anything Parameter fault Correction of setting as shown in the manual...

Страница 11: ...omestic waste Batteries and rechargeable batteries are not allowed in domestic waste and have to be disposed fully discharged or with insulated poles Otherwise there s a danger of an electrical short Please ask at your local administration for an environmentally correct disposal 6 4 Warranty A one year warranty period from date of purchase will be guaranteed for either repair or replacement once i...

Страница 12: ... EU regarding electromagnetic compatibility RoHS Directive 2011 65 EG as appropriate amended This declaration loses their validity in case modifications on the scales are made without manufactures approval Hamburg January 2019 ADE Germany GmbH Co KG Hammer Steindamm 27 29 22089 Hamburg Germany 7 Declaration of Conformity Manufacturer ADE Germany GmbH Co KG Hammer Steindamm 27 29 D 22089 Hamburg Ge...

Страница 13: ...115 mm Dead Weight 4 1 kg 3 8 kg Connecting cable 1 00 m 2 00 m Operating Temperature 10 C to 40 C 10 C to 40 C Internal Protection IP65 IP68 Power Supply 230V mains 230V mains and rechargeable battery 6 3V 6V4Ah refitting with rechargeable battery not possible Display Large LCD Display with second display on backside height of digit 20mm Casing Stable stainless steel housing Weighing platform sta...

Страница 14: ...13 KWE 190107 Rev008 UM en ...

Страница 15: ...KWE 190107 Rev008 UM en 14 ...

Страница 16: ...15 KWE 190107 Rev008 UM en Händler ...

Отзывы: