9
KWE-190107-Rev008-UM-en
3.8
Printer setting
Please start as described in the beginning of chapter 3 to enter this menu and „Prt_“ is shown
in the display.
Setting has to be “Prt_0”, because scale has no interface for connecting PC or printer.
4.
Display Symbols
SP – x.x
Version number
„xxxxxx“
Software number
“CAPxx.0”
Capacity / Maximum load of the scale
„nnnnnn“
Scale is overloaded. The current weighing is higher than the valid
capacity.
“------“
Data is unstable
“LLLLLL”
Output of load cell is lower.
“HHHHHH”
Output of load cell is higher or turn on zero is higher than range.
- Lo -
Low battery – Recharge the rechargeable battery immediately.
FULL
Rechargeable battery has been charged fully.
“_Inlt_”
Faulty weighing
5.
Display Symbols
Fault
Cause
Solution
Faulty weighing result
Parameter of load cells have
changed
Re-calibration
Display shows „HHHHHHHH“ Over zero range
Re-calibration
Load cell or PCB damaged
Display shows „LLLLLLLL“
Load cell or PCB damaged
Rear display doesn’t show
anything
Parameter fault
Correction of setting as
shown in the manual
Load cell damaged
Rechargeable battery is not
loading although loading is
shown
Rechargeable battery is
damaged
Replace rechargeable battery
Rechargeable battery is not
loading and no loading is
shown
Charging unit for
rechargeable battery is
damaged
Quickly discharging of
rechargeable battery
Old rechargeable battery
Replace rechargeable battery
Display shows „_InIt_“
Faulty weighing
Re-start the scale or Re-
calibration