Error messages
Failure
Description
Remedy
- - - - -
Overload
Remove the weight from the scale.
If the problem persists, contact your dealer.
Err 1
Incorrect date entry
Enter the date with correct and logical values in the
format yy:mm:dd.
Err 2
Incorrect time entry
Enter the time with correct and logical values in the
format hh:mm:ss.
Err 4
Incorrect zeroing
The scale was out of the normal zero range when
the scale was switched on or when the “Zero” key
was pressed.
Take the weight off the scale and repeat the
process.
If necessary, press the “TARE” key to reset the
display.
If the problem persists, contact your dealer.
Err 6
A/D outside the
display range
The measured values of the A/D converter are
outside the normal display range.
Remove the overloading weight from the scale and
make sure that the weighing platform is correctly
positioned.
If the values remain unchanged, this indicates that
the measuring cell or the electronics could be
damaged.
If the problem persists, contact your dealer.
Storage and Transport Conditions
Keep all original packaging materials for a possible return of the scale to avoid damage in transit.
Disconnect all cables before transport and switch off the scale to avoid damage.
Screw in the transport lock for transport.
Disposal
Waste electrical equipment do not belong in household waste. Devices
with this marking may not be disposed of as residual waste, but must
be recycled.
Remove (if possible) all batteries and rechargeable batteries from the
devices and send them to the battery disposal.
Make sure that only depleted batteries or batteries with insulated poles
are disposed of so that there is no short circuit!