41
the range. The sequence of calibration is to calibrate the lower limit first, then zero point, and finally the upper
limit. After the upper limit is finished, return to the calibration setting interface.
4. Cancel calibration: In the pressure calibration setting interface, click the main display area to the cancel calibration
interface, and click OK to cancel the calibration data.
5. Cancel zero: In the pressure calibration setting interface, click the main display area to the cancel zero interface,
click OK to cancel the zero.
Figure23 Pressure calibration interface