43
calibration interface, the main display area displays the measured value of barometric pressure, the left side of the sub
display area indicates the current calibration point, and the right side indicates the standard value, it can be clicked to
edit. Click OK to enter the next point, and finally submit the calibration data to complete the calibration.
Figure25 Barometric calibration selection interface
3.5.4
(4~20) mA Calibration (only available when using (4~20) mA module)
When the (4~20) mA module is enabled, the module can be calibrated. Two-point calibration and cancel calibration
are provided. The main display area displays the output value, the left side of the sub display area displays the
calibration point, and the right area displays the entered standard value. Click OK to go to the next point.