45
3.7.5
Overpressure Record Information
The title bar displays OVER P, Press the main display area to switch to the next overpressure record information, the
main display area displays the total number of overpressure records, the current serial number, the sub display area
displays the overpressure value, short press the sub display area to switch to display the date and time of the
overpressure record.
3.7.6
Touch Sound Settings
The title bar displays TOUCH, Press the main display area to edit the touch sound settings, it supports touch sound
on and off.
3.8
Malfunction Information
Press the Check icon in the menu bar to enter the equipment malfunction interface. When there is nothing wrong, it
will display NORMAL. When there is malfunction, the corresponding fault code will be displayed. Click the main display
area to switch to display different fault codes.
Table10 Malfunction information list
Error codes
Description
Solution
101
Ferroelectric self-test failed
Restore to factory settings
102
Pressure module init failed
Check the pressure module
connections
103
Barometric module init failed
Please contact the manufacturer
104
Barometric module calibration init failed
Please contact the manufacturer
Содержание AD685Ex
Страница 1: ...685 685Ex Advanced Digital Pressure Gauge...
Страница 2: ...ADT685 685EX Advanced Digital Pressure Gauge User Manual Version 2103V01 Additel Corporation...
Страница 4: ......
Страница 8: ......
Страница 19: ...11 1 5 Basic Structure Figure 1 Basic Structure...
Страница 63: ...55...