115
8. System maintenance
8.1 Device information view
On the system setting interface, select the product information. For details, see Para. 4.9 Product information.
When the external pressure module is online, it can also view its related information.
8.2 Diagnostic information
In case of any abnormalities of the calibrator, there will be a corresponding prompt message, and the message
notification center icon on the top status bar will turn red and flash for alarm.
8.3 Check the sealing performance of the calibrator
The sealing performance of the calibrator has a crucial impact on its control speed and control pressure stability. Slow
pressurization speed, long stabilization time of the control output, and too much fluctuation of the output pressure in the
process may all be caused by poor sealing. The method of checking the sealing performance of the calibrator is as
follows:
Use a plug (standard fitting) to block the outlet (OUTLET).
Go to the diagnostic test under maintenance, refer to Para. 4.5.2, select the low/high pressure module pressure
leakage and emptying test, and check the results after the self-test of the test is completed.