Table 14 The settings of the differential temperature module
Subject
Valid Value
Description
Enable
Enable /disable
Enable or disable the differential temperature module. When the
differential temperature module is enabled, the differential
temperature module channels will be displayed in the related
functions channel list.
Resolution
4/5/6
The resolution of differential temperature
Range
Numbers
Differential module measuring range
Calculate type
CH1 - CH2 or CH2 - CH1
For the composite mode of differential temperature module, CH1
and CH2 represent two temperature measurement channels at
the top of the device
Real time data
Real time differential temperature value
Read-only. When the temperature difference module is enabled
and the two channels are measuring normally, the real-time
temperature difference value will be displayed,
Otherwise, the real-time differential temperature value is
displayed as ---------
Align offset
The difference between two channels
calculated when aligning the channels
Read only. When performing channel aligning, it will be changed
and saved to the temperature difference between channels
CH1
Module real time values and units
Read only
CH2
Module real time values and units
Read only
Содержание 282
Страница 1: ...282 Dual Channel Reference Thermometer Readout...
Страница 2: ......
Страница 3: ...ADT282 Dual Channel Reference Thermometer Readout User Manual Version 2204V01 Additel Corporation...
Страница 20: ...1 4 Basic Structure Figure 1 Basic Structure...
Страница 44: ......