24
two modules are online. Select the proper module in the Channel settings to measure the pressure. Click the pressure
unit above the pressure measurement value to switch the pressure unit in the pop-up unit selection interface.
Click the menu icon
in the pressure measurement channel and click the "Settings" menu in the pop-up menu to set
the pressure type, resolution, stability, tare, and other functions of the current pressure measurement channel.
Table 6 Pressure measurement settings
Subject
Valid Value
Description
Resolution
4/5/6
Resolution of current measurement channel
Pressure type
GP/ AP
Current measurement channel pressure type
Measurement frequency
1~10Hz
Data acquisition frequency of the current measurement
channel
Stability enable
Enable/disable
Whether to turn on the stability judgment function
Stability value
Pressure value or percentage of
range
A value for judging whether the pressure is stable. If the
measured value fluctuates less than this value within the
stable time, the calibrator will judge as stable
Stabilization time
Number
Time to judge the pressure stability
Tare enable
Enable/disable
Enable or disable tare function
Tare value
Pressure value
Correction value of the tare function