36
6. Data logger
The most important function of the device is the data logger. Click the menu icon
at the top right of the main
interface and click "Start logging" to enter the logging configuration interface. Here it is available to set the interval for
each logging point and the total number & time of loggings. After finishing the configuration, click OK to start.
During the process, the measured values of all currently selected channels and the time are recorded and saved in a
file. After the logging starts, the progress bar at the bottom and the percentage will show the progress, and the time
beside the progress bar will show the remaining time.
Click the stop icon
on the upper right to terminate the current logging, and you can choose whether to store the data
that has been recorded during the process.
6.1 Display mode
The ADT260Ex supports single-channel mode and multi-channel list mode, click the icon
at the left of the progress
bar to switch the display mode.
6.1.1 Single-channel mode
Under this mode, both the curve and data will be displayed at the same time, as shown in figure 11. In case of several
channels need to be recorded, users can switch to the curve and data of other channels manually or automatically.