4. Position the Stethoscope:

Position the chestpiece in the antecubital space below 

the cuff, distal to the brachium. Do not place the chestpiece underneath the cuff, as 
this impedes accurate measurement. Use the bell side of a combination stethoscope for 
clearest detection of the low pitched Korotkoff (pulse) sounds.

5. Systolic Assist Function and Pulse Rate Display:

Your digital e-sphyg

TM

gauge is 

equipped with a systolic assist function to help you more easily determine the onset of
systolic blood pressure. As you deflate the cuff, a heart symbol will flash when the 
pressure is within about 20 mmHg of systolic pressure. Once you see the flashing 
symbol, pay close attention for the start of phase 1 Korotkoff sounds. The digital 
gauge will also measure the patient’s pulse while you are measuring blood pressure, 
and display this information on the LCD screen once the cuff has been deflated.

Note:

Systolic assist and pulse display features are NOT motion tolerant. Excessive 

movement may interfere with accurate display of these features. Excessive movement 
will NOT interfere with accurate determination of blood pressure using the e-sphyg

TM

.

6. Deflate the cuff:

Open the valve to deflate the cuff gradually at a rate of 2-5 mmHg 

per second (2-3 is preferred). The display will show the average deflation rate in the 
lower right corner of the LCD screen.

Caution:

Excessive deflation rates (in excess of 7mmHg) will lead to inaccurate 

measurement and will make it difficult to pinpoint the readings on the display.

7. Measurement:

Record the onset of Korotkoff sounds as the systolic pressure, and 

the disappearance of these sounds as diastolic pressure. (Some healthcare 
professionals recommend recording diastolic 1 and diastolic 2. Diastolic 1 occurs at 
phase 4).

After measurement is completed, open valve fully to release any remaining air 
in the cuff. The digital gauge will now display the patient’s pulse. Remove cuff.

Backlight Control

The default setting on your digital gauge is to have the back light on once the 
pressure is inflated over 30 mmHg, and throughout the measurement process. 
To switch the backlight off at any time, press the “       ” button on the left side. 
You will need to deactivate the backlight each time you power on the instrument.  

Assembling your Sphygmomanometer

Your inflation system is pre-assembled and ready for use. You will need to simply
connect your digital gauge to the complete inflation system by inserting the twist lock
connector (pre-assembled on a bladder tube) into the air port at the bottom of the digital
gauge, twist to lock.

If you will be hanging the gauge from the cuff gauge holder, it is recommended that the
bladder tubing connecting the gauge be trimmed to a length of 6” to 9” for convenience.
(Note: Be sure to retain the twist lock connector needed to attach the digital gauge to the
tubing.)

Measurement Procedure

(Auscultatory Method)

1. Patient Position

: The patient should sit or lie comfortably. The arm should be fully 

supported on a flat surface at heart level. (If the arm’s position varies, or is not level  
with the heart, measurement values obtained will not be consistent with the patient’s 
true blood pressure.) 

2. Apply the cuff:

To attach the digital gauge, plug the connector into the gauge and 

twist clockwise to ensure a secure connection. The e-sphyg

TM

gauge can be either 

clipped on the gauge hang tab, or held in your hand during use. ADCUFF

TM

nylon cuffs 

are specially designed to promote the accurate determination of blood pressure. Index 
and range markings ensure use of the correct cuff size. The artery mark indicates 
proper cuff positioning.

Place the cuff over the bare upper arm with the artery mark positioned directly over 
the brachial artery. The bottom edge of the cuff should be positioned approximately 
one inch above the antecubital fold. Wrap the end of the cuff not containing the 
bladder around the arm snugly and smoothly and engage adhesive strips. To verify a 
correct fit, check that the Index Line falls between the two Range Lines.

3. Activate the digital aneroid and inflate the cuff:

Press the “       ” On/Off button to 

activate your digital gauge. After zeroing, your digital gauge is ready to measure. Close 
the ADFLOW

TM

valve by turning thumbscrew clockwise. Palpate the radial artery while 

inflating the cuff. Be sure to inflate cuff quickly by squeezing bulb rapidly. Inflate cuff 
20-30 mmHg above the point at which the radial pulse disappears.

Содержание E-Sphyg 7002

Страница 1: ...m email info adctoday com 2007 ADC All rights reserved No one is permitted to reproduce or duplicate in any form this manual or any part thereof without the expressed written permission of ADC ADC ass...

Страница 2: ...he finest sphygmomanometers that money can buy With proper care and maintenance your ADC blood pressure instrument is sure to provide you with many years of depend able service Please read the followi...

Страница 3: ...normal use and service as follows 1 Warranty service extends to the original retail purchaser only and commences with the date of delivery 2 The entire sphygmomanometer is warranted for one year 3 The...

Страница 4: ...and accuracy of your new digital e sphygTM gauge Using state of the art technology your e sphygTM will provide digital accuracy with tank like reliability Intended Use ADC aneroid digital aneroid and...

Страница 5: ...A warning statement in this manual identifies a condition or practice which if not corrected or discontinued immediately could lead to patient injury illness or death NOTE Systolic Assist and pulse m...

Страница 6: ...utton On Off Zeroing Button Artery Mark Air Tube and Connector Air Release Valve Inflation Bulb Splash Resistant Housing Extra Large LCD Display Error Message Description Solution ERROR DISPLAYS and T...

Страница 7: ...nician is qualified to perform this calibration service Do not attempt to recalibrate the gauge yourself In addition to the visual inspection indicated above we also recommend you perform the followin...

Страница 8: ...the backlight off at any time press the button on the left side You will need to deactivate the backlight each time you power on the instrument Assembling your Sphygmomanometer Your inflation system i...