Introduction

Blood pressure measurements determined with this device are equivalent to those obtained by
a trained observer using the auscultatory method, within the limits prescribed by the American
National Standards Institute, ANSI/AAMI SP10: 2002, Manual, Electronic, or Automated
Sphygmomanometers.

Your new ADC

®

e-sphyg

TM

gauge provides you with all of the functionality of a traditional

aneroid gauge, with none of the problems associated with these mechanical devices.
Using state of the art technology, your e-sphyg

TM

gauge will provide you with performance

and reliability exceeding the most popular aneroid models. While traditional mechanical
gauges can be easily knocked out of calibration when dropped, your digital gauge will
remain accurate even when subjected to rough treatment. With its helpful on-screen cali-
bration reminder, you will be able to use your new digital gauge with confidence and with
the knowledge that it will remain calibrated beyond the lifespan of a more traditional
mechanical gauge.  

Your digital e-sphyg

TM

aneroid also has several features not available in a more traditional

mechanical gauge. The systolic assist function will give you a visual alert when you are
approaching the patient’s estimated systolic pressure, allowing you to focus your attention
to detect the beginnings of phase 1 Korotkoff sounds. The digital gauge will also monitor
the patient’s pulse during the measurement process, and display this information
onscreen for you to record.

We hope you enjoy the convenience and accuracy of your new digital e-sphyg

TM

gauge.

Using state-of-the-art technology, your e-sphyg

TM

will provide digital accuracy with tank-like

reliability.

Intended Use

ADC

®

aneroid, digital aneroid, and mercurial sphygmomanometers are used by professional

healthcare providers and individuals trained in auscultatory blood pressure technique to
determine systolic and diastolic pressure in humans.

Specifications

Power Source:

Two 1.5V DC LR03 (AAA) Alkaline Batteries

MEASUREMENT RANGE

Pressure: 0-299 

mmHg

Pulse: 

30-240 beats/min Accuracy

Pressure:

±3 mmHg

Pulse:

Within ±5% of reading

Pressure Sensor:

Semi-conductor

Inflation:

Manual

Deflation:

Manual Air Release Valve

Auto-shut-off:

1 min. after last key operation

Backlight auto-shut-off:

15 sec. after last key operation

OPERATION ENVIRONMENT

Temperature:

32~115°F (0~46°C)

Humidity:

85% Relative Humidity max

STORAGE ENVIRONMENT

Temperature:

-4~131°F (-20~55°C)

Humidity:

90% Relative Humidity max (non condensing)

Dimensions:

3”(L) X 2

3/8

”(W) X 1

1/8

”(H)

Weight:

3.14 oz. (G.W.) (w/o Batteries) Type BF

Battery Life:

1500 readings with back light on
3000 readings with back light off

Device and cuff are designed to provide special protection against electrical shocks.
Specifications are subject to change without notice. 

Standards

ADC’s digital aneroid sphygmomanometer is accurate to +/- 3 mmHg
and conforms to applicable sections of the following international standards:

EV council directive 93/42/EEC

EMC directive 89/336/EEC

EN 1060-1: 1995, Non-invasive sphygmomanometers - Part 1:

General Requirements

EN 1060-3: 1997, Non-invasive sphygmomanometers - Part 3:

supplementary requirements for electro-mechanical blood pressure measuring systems

ANSI / AAMI SP10: 2002, Manual, Electronic, or Automated Sphygmomanometers

Содержание E-Sphyg 7002

Страница 1: ...m email info adctoday com 2007 ADC All rights reserved No one is permitted to reproduce or duplicate in any form this manual or any part thereof without the expressed written permission of ADC ADC ass...

Страница 2: ...he finest sphygmomanometers that money can buy With proper care and maintenance your ADC blood pressure instrument is sure to provide you with many years of depend able service Please read the followi...

Страница 3: ...normal use and service as follows 1 Warranty service extends to the original retail purchaser only and commences with the date of delivery 2 The entire sphygmomanometer is warranted for one year 3 The...

Страница 4: ...and accuracy of your new digital e sphygTM gauge Using state of the art technology your e sphygTM will provide digital accuracy with tank like reliability Intended Use ADC aneroid digital aneroid and...

Страница 5: ...A warning statement in this manual identifies a condition or practice which if not corrected or discontinued immediately could lead to patient injury illness or death NOTE Systolic Assist and pulse m...

Страница 6: ...utton On Off Zeroing Button Artery Mark Air Tube and Connector Air Release Valve Inflation Bulb Splash Resistant Housing Extra Large LCD Display Error Message Description Solution ERROR DISPLAYS and T...

Страница 7: ...nician is qualified to perform this calibration service Do not attempt to recalibrate the gauge yourself In addition to the visual inspection indicated above we also recommend you perform the followin...

Страница 8: ...the backlight off at any time press the button on the left side You will need to deactivate the backlight each time you power on the instrument Assembling your Sphygmomanometer Your inflation system i...

Отзывы: