• The maximum charging time for NiMH rechargeable batteries is
limited to <650 minutes (10.8 hours). The maximum charging time for
Li-ion rechargeable batteries is limited to <480 minutes (8 hours).
• When the rechargeable battery is fully charged (green LED on), the
unit will switch to maintenance charge.
• The temperature inside the charger base is monitored. Do not keep
in sunlight, or close proximity of a radiator, as over heating may
cause premature actuation of the automatic switch-off mechanism
shutting the charger off before completing a full charge.
Caution:
• When the handle is removed from the charging chamber for a few
seconds while the rechargeable battery is fully charged (green LED on)
without switching on, initially the yellow charging LED will be switched
on when reinserting the unused rechargeable
battery. However, this is not due to
malfunction of the charging base but the
described “rechargeable battery full” sensor.
• When the rechargeable battery is full, no
negative voltage differential can be recorded.
The rechargeable battery must first heat up in
order to signal “rechargeable battery fully charged.”
• The time prior to the green “rechargeable
battery fully charged” LED switching on
again may vary considerably (15 minutes
or longer). Additional temperature sensing
and a timer, however, ensure that the
charging process is carefully monitored at
any time.
We recommend keeping the instrument in the
charging chambers after use to ensure that instruments are always ready.
11. TECHNICAL SPECIFICATIONS
Model:
ADC
Rechargeable Battery Charger Base
Display System:
LEDs Connection
Main Connection
(See Note on Bottom Label)
Power Adapter Input:
100 V-240 V AC / 50-60 Hz/0.6 A
Power Adapter Output:
9 V DC/2 A/18 W
Classification:
Applied part type B
Operating Conditions:
32°F to 104°F (0°C to 40°C)
10% to 85% relative humidity
Storage Temperature:
23°F to 122°F (-5°C to 50°C)
10% to 85% relative humidity
Air Pressure:
700 to 1050 hPa
Dimensions:
7.3” x 3.4” x 2.9” (186 mm x 87 mm x 74 mm)
Weight:
Approx. 1.1 lbs (500 g) without packing
12. MAINTENANCE
These instruments and their accessories do not require any specific
maintenance. Should an instrument have to be examined for any specific
reason whatsoever, please return it to ADC.
13. ELECTROMAGNETIC COMPATIBILITY (EMC)
Follow the instructions during the installation and operation of the device.
To avoid electromagnetic interference with the operation of the device, do not
use the device at the same time as other electronic devices.
To avoid electromagnetic interference with the operation of the device, do not
use or stack the device near, on or under other electronic devices.
Do not use the device in the same room as other electronic devices, (e.g. life-
supporting devices that have a significant impact on the life of the patient and
treatment outcomes) or other measuring instruments or treatment devices that
use little electrical power.
Do not use cables or accessories that are not specified for this device, as this
can increase the emission of electromagnetic waves from the device and reduce
the device’s electromagnetic immunity.