Warning:
The device must not be operated in rooms in which flammable
mixtures or mixtures of pharmaceuticals and air or oxygen or nitrous oxide are
present, e.g. operating theatres.
Warning:
Do not use in a magnetic resonance environment.
Warning:
There is a danger of life-threatening electrical shock. Unplug the
instrument before cleaning or when disinfecting.
Warning:
The device contains sensitive electronic components. Avoid strong
electrical or electromagnetic fields in the direct vicinity of the device (e.g.,
mobile telephones, microwave ovens). These can lead to temporary impairment
of the ADC Diagnostix Desk Charging Base.
Warning:
This device must be used in a controlled environment. The device must
not be exposed to harsh environmental conditions.
Warning:
Do not use batteries or electrical cords other than those included with
this product or replacement parts supplied by the manufacturer.
Caution:
Old electronic devices must be disposed of in accordance with the
institutional guidelines for the disposal of expired devices.
6. START-UP AND OPERATION:
The ADC Diagnostix Desk Charging Base is suitable for charging of 2.5V or
3.5V handles.
a) Preparation prior to charging
• Ensure that the charging base is
positioned on a level surface. Do not
position unit in the direct vicinity of a
radiator. Protect against direct sunlight.
• Fluids must not penetrate into the interior
of the enclosure. Do not use fluids in close
proximity of the charging base.
Guidance and manufacturer‘s declaration - electromagnetic immunity
The charging station is intended for use in the electromagnetic environment specified below.
The customer or user of the charging station should ensure that it is used in such an environment.
Hinweis UT is the AC mains voltage prior to application of the test level.
Immunity testing
Electrostatic discharge (ESD)
IEC 61000-4-2
Fast transient electrical disturbances /
bursts IEC 61000-4-4
Con: ± 8 kV
Air: ± 15 kV
5/50 ns, 100 kHz;
±2 kV
5/50 ns, 100 kHz;
±2 kV
Con: ± 8 kV
Air: ± 15 kV
Floors should be made of wood, concrete or
ceramic tiles. If the floor is covered with synthetic
material, the relative humidity must be at least
30%.
The quality of the supply voltage should be that of a
typical business or hospital environment
Surge
IEC 61000-4-5
1,2/50 (8/20) μs
LtL: ±1,0 kV
LtG: ±2,0 kV
1,2/50 (8/20) μs
LtL: ±1,0 kV
LtG: ±2,0 kV
The quality of the supply voltage should be that of a
typical business or hospital environment.
Voltage dips, short interruptions and
voltage fluctuations on power supply
input lines
IEC 61000-4-11
Power frequency (50Hz/60Hz)
magnetic field IEC 61000-4-8
0% UT for 0.5
cycle
(1 phase)
0% UT for 1 cycle
70% UT for 25/30
cycles (50/60 Hz)
0% UT for 250/300
cycles (50/60 Hz)
30 A/m
50 Hz
30 A/m
50 Hz
0% UT for 250/300
cycles (50/60 Hz)
0% UT for 0.5
cycle
(1 phase)
0% UT for 1 cycle
70% UT for 25/30
cycles (50/60 Hz)
The quality of the supply voltage should be that of a
typical business or hospital environment.
Mains frequency magnetic fields should be at a
level characteristic of a typical location in a
typical commercial hospital environment.
IEC 60601 test level
Compliance level
Electromagnetic environment guidance
5