4. SYMBOLS
5. WARNINGS AND PRECAUTIONS
This ADC Diagnostix Desk Charging Base has been manufactured to the
highest quality standards and is subject to rigorous quality control. Read
these instructions for use carefully before putting the unit into operation and
keep them in a safe place.
If you should have any questions, call our toll-free number or visit our
website. Our address can be found on the last page of this booklet.
Please note that instruments described in these instructions for use are to be
used by suitably trained personnel only. The safe functioning of this
instrument is guaranteed only when original parts and accessories from ADC
are used.
Read operating instructions
Medical device
Type BF Equipment
Protection Class II Unit
Warning!
The general warning
sign indicates a possibly dangerous
situation that can lead to serious
injuries.
Caution!
Important note in these
instructions. The caution symbol
indicates a potentially dangerous
situation that can lead to minor or
moderate injuries. It can also be
used to warn of unsafe practices
Direct current
Alternating current
Date of Manufacture
YYYY-MM-DD / (year-month-day)
Manufacturer
Manufacturer serial number
Lot number
Reference number
Temperature for transport and
storage conditions
Relative humidity for transport and
storage conditions
Air pressure for transport and storage
Air pressure for ambient operation
CE marking
Discard according to national
regulations or EU directives
Non-ionizing radiation
Caution:
Federal law restricts this
device to sale by or on the order of a
physician (licensed physician).
Not made with natural rubber latex
Phthalate free
SN
Symbol Definition
Symbol Definition
Guidance and manufacturer‘s declaration - electromagnetic immunity
The charging station is intended for use in the electromagnetic environment specified below.
The customer or user of the charging station should ensure that it is used in such an environment.
NOTE 1:
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2:
These guidelines may not apply in all situations. The electromagnetic propagation is affected by absorption and reflection by
structures, objects and people.
a.)
Field strengths from fixed transmitters, such as base stations for radio (mobile/cordless phones) and land mobile radios, amateur radio, AM and
FM radio broadcasts and television broadcasts, cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location where the charging
station is used exceeds the RF compliance level specified above, the charging station should be monitored to verify normal operation. If abnormal
performance is observed, additional measures may be required, such as re-orienting or relocating the charging station.
b.)
With a frequency range over 150 kHz to 80 MHz, the field strengths should be less than 3 V/m.
Immunity testing
Conducted RF
IEC 61000-4-6
Fast transient electrical disturbances /
bursts IEC 61000-4-4
Proximity fields from RF wireless
communications equipment
Pass
Pass
Pass
3 V/m
80 Mhz to
2,7 GHz
3 V/m
Pass
Portable and mobile RF communications equipment
should be used no closer to any part of the
charging station, including the cables, than the
recommended distance calculated using the
equation applicable to the frequency of the
transmitter.
Recommended separation distance
d= 1.2√P 150 KHz to 80 MHz
d= 1.2√P 80 MHz to 800 MHz
d= 2.3√P 800 MHz to 2,7 GHz
Where P is the maximum output power of the
transmitter in watts (W) according to the transmitter
manufacturer and d is
the recommended distance
in meters (m). Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey, should be below the
compliance level in each frequency range. In the
vicinity of devices marked with the following symbol,
interference may occur:
Where P is the maximum output power of the
transmitter in watts (W) according to the transmitter
manufacturer and the recommended distance is
indicated in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey, should be below the compliance level in
each frequency range.
IEC 60601 test level
Compliance level
Electromagnetic environment guidance