Preventative Maintenance
System Self Checks
The ADView 2 performs a range of system checks during normal operation. If the device detects a problem, it will
display an error code.Replaceable Parts
On a routine basis, inspect the monitor, cuffs, and hoses for cracks, fraying, or kinks. Immediately replace any
damaged part. Refer to the list of Accessories & Replacement Parts in this guide.
Replacing and Disposal of the Battery
CAUTION: Fire, explosion and severe burn hazard. Replace only with ADC part number 9005BAT.
When the battery no longer charges or it needs excessive recharging, it may need to be replaced. See the “Installing
Battery” section above for proper installation and precautions to be taken when installing the rechargeable battery.
Please dispose of the old battery per the instructions below.
Battery Disposal
The ADView 2 device contains a lithium-ion battery that contains materials which may be hazardous to human
health. Do NOT dispose of battery in domestic waste! Instead, please dispose of in an environmentally responsible
way, or return the battery to ADC. A prepaid return label can be obtained. Please see our website for more
information.
Product Disposal
Do not dispose of this product as unsorted municipal waste. Prepare this product for reuse or separate
collection as specified by Directive 2002/96/EC of the European Parliament and the Council of the European Union on
Waste Electronic and Electrical Equipment (WEEE).
Cuff Disposal
Do not return used cuffs. Used blood pressure cuffs may be contaminated medical waste and should be dealt
with in accordance to your local regulations.
Routine Calibration
Have the ADView 2 blood pressure function checked every two years to verify the accuracy of the pressure
transducers and indicators. Please have this done by a biomedical technician or ADC Customer Service.
CAUTION:
Calibration should be done by a biomedical technician or other person familiar with the ADView 2
device.
If a biomedical technician performs this function, they should contact ADC for instructions to access “Verify
Calibration.” Instructions are also available in the ADView 2 Advanced Configuration manual (ADC Part 93-9005AC-
00).