ADView 2 User Manual | 36
93-9005-00
rev 4
September 1, 2016
and Check Mark Icon
will start flashing. Press the Selection Knob to send the data to the EMR. If the
measurements are delivered to the EMR successfully, you will see the EMR Message Icon
and the Check Mark
Icon
flash four times.
If the EMR has a problem receiving the measurements, the EMR Message icon
and X icon
will flash eight
times and you will hear 4 short beeps. You will see a message indicating the nature of the problem, such as:
“Rejected” or “Rejected for an error.” This message occurs when the device is connected to the network but the EMR
rejects the message. A corresponding message will appear in the Patient ID field during a three-second period when
the Message Icon
and X icon
flash on and off.
“No response (time out).” This message occurs when the device is connected to the network but the EMR is not
responding. It also can occur if network connectivity is lost. If there is a network connection, the ADView 2 will
continue to try to transmit the message every 10 seconds. A message in the Patient ID field will alert you that another
attempt to send the message will be made in 10 seconds. If network connectivity is lost, you will see a corresponding
message in the Patient ID field, and the device will continue trying to re-establish a connection and send the
measurements at intervals set during EMR setup. If problems persist, you may wish to contact your facility’s
information technology department. As a backup, the patient’s measurements are saved in the device memory when
the memory button was pressed, and can be retrieved using the procedure detailed in the beginning of Section 9.
Note: After the memory button is pressed, all values to be stored to memory will flash off and on twice, along with the
check mark icon, memory icon and memory location identifier. After flashing, the display will revert to null values for
the patient ID and all measurement data, and the check mark icon will flash off and on twice more.
10. Taking Care of Your ADView 2
Cleaning
CAUTION:
The ADView 2
is not sterilizable. Do not immerse the monitor in any fluid or attempt to clean with
any liquid detergents, cleaning agents, or solvents.
Dampen a soft cloth with mild medical grade disinfectant and wipe the device to remove surface dust and dirt. Dry
surface thoroughly before use.
SpO
2
Sensors and Thermometers
CAUTION:
Never immerse sensors, clips or thermometers in fluids. Do not pour or spray any liquids onto the
sensor or thermometers. Caustic or abrasive cleaners will cause permanent damage. Do not open the case of the
Sp02 sensor finger clip sensor more than 45° or the case will be damaged.
Clean the Sp02 sensor and thermometers with a soft cloth dampened with a mild medical grade disinfectant or
isopropyl alcohol. Remove all tape residues. Allow the sensor and thermometer to dry thoroughly before reusing.