Other possible errors and their solutions

If problems occur when using this device, the following points should be checked.

NOTE:

Blood pressure is subject to fluctuations even in healthy people. 

Please remember that

comparable blood pressure measurements always require the same time and the same
conditions!

These are normally quiet conditions. If you follow the procedure described earlier and still get

fluctuations of blood pressure of greater than 15 mmHg and/or you repeatedly hear irregular pulse tones,
consult your doctor.

Attention!

Should any technical malfunctions arise in the blood pressure monitor, please contact the dealer where you
bought the device or ADC. 

Never attempt to repair the instrument yourself!

Any unauthorized

Error No.

ERR 1

ERR 2

ERR 3

Possible Cause(s)/Solutions

Pulse could not be detected. Please ensure the cuff is properly
placed on the inside of your wrist and repeat the measurement.

Unnatural pressure impulses influence the measurement
result. Reason: the arm was moved during the measurement.

If inflation of the cuff takes too long, the cuff is not correctly
seated. Reposition the cuff and repeat the measurement.

ERR 5

HI

LO

The difference between systolic and diastolic is excessive.
Measure again carefully following proper procedures. Consult
your doctor if you still get unusual results of measurements.

The pressure in the cuff is too high (over 300mmHg). Relax for
5 minutes and repeat the measurement.

The pulse is too low (less than 40 beats per minute). Relax for
5 minutes and repeat the measurement.

Malfunction

The display remains blank when
the device is switched on though
the batteries are in place.

The device frequently fails to
measure, or the values
measured are too low or high.

Every measurement results in
different values, although the
device functions normally and
normal values are displayed.

Blood pressure values differ from
those measured at the doctor’s.

Remedy

1. Check battery installation/ polarity.
2. Remove the batteries and if the display is
unusual, then exchange them for new ones.

1. Fit the cuff correctly on the wrist.
2. Before starting measurement make sure
that clothing is not exerting pressure on the
arm. Take articles of clothing off if necessary.
Measure blood pressure again in complete
peace and quiet.

Refer to 

Section 6.2 Common

Sources of Error.

Refer to 

Section 3.7. Why Measure

Blood Pressure at Home?

Содержание Advantage 6015N

Страница 1: ...15N 6016N Digital Wrist Blood Pressure Monitor INSTRUCTION MANUAL PLEASE NOTE THIS MEDICAL INSTRUMENT MUST BE USED ACCORDING TO INSTRUCTIONS TO ENSURE ACCURATE READINGS Questions Call ADC toll free at 1 800 232 2670 ...

Страница 2: ... place Please contact your doctor for further questions on the subject of blood pressure and its measurement To learn more visit our website at www adctoday com Remember Only a healthcare professional is qualified to interpret blood pressure measurements This device is NOT intended to replace regular medical checkups It is recommended that your physician review your procedure for using this device...

Страница 3: ...edback from the nervous system To adjust blood pressure the strength and frequency of the heart Pulse as well as the width of circulatory blood vessels is altered Blood vessel width is effected by fine muscles in the blood vessel walls Blood pressure is traditionally measured in millimeters of mercury mmHg It is recorded as systolic diastolic For example a systolic of 120 and diastolic of 80 would...

Страница 4: ...ecific causes are unknown although the condition is often linked with family history and lifestyle This is referred to as essential hypertension In the remaining cases high blood pressure is a symptom of an underlying often treatable condition which if corrected may normalize blood pressure This less common type is known as secondary hypertension Hypertension if left untreated may contribute to ki...

Страница 5: ...r is a system consisting of an inflatable bladder contained within a cuff inflation bulb with air control valve and pressure measuring manometer gauge The gauge may be mechanical electronic or mercurial The cuff is wrapped around the limb and inflated to constrict blood flow to the artery As pressure is released from the cuff through the deflation valve blood flow returns to the artery producing p...

Страница 6: ...e batteries from the device Use AAA Long Life or Alkaline 1 5V batteries Do not use rechargeable batteries Functional check Press the MEMORY button down to test all the display elements When functioning correctly many icons will appear 5 2 Setting the Time and Date This blood pressure monitor incorporates an integrated clock with date display After new batteries have been inserted the device will ...

Страница 7: ...t 1 then only one measurement will be taken If you select 3 the unit will inflate and deflate three times resulting in one final average 5 5 Using ADC Average Mode Technology 6016N only In MAM Advantage Averaging mode three measurements are automatically taken in 1 succession and the result then automatically analyzed and displayed Because your blood pressure constantly fluctuates a result determi...

Страница 8: ...uff Remove all objects and jewelry e g wristwatch from the wrist in use Put the 1 cuff over the wrist with the monitor face on the inside of the wrist The buttons should be on the bottom The distance between the cuff and the hand should be approx 10 mm 1 2 2 Secure the cuff with the hook and loop fastener so that it lies comfortably and 3 not too tight No space should remain between the cuff and t...

Страница 9: ...tant that you be relaxed remain still and do not talk during measurements NOTE We recommend contacting your physician if you see this indicator frequently To prolong the batteries life the device switches off automatically if no button is pressed for 2 minutes Otherwise you can switch it off by pressing the ON OFF button 7 SOFTWARE FUNCTIONS 6016N ONLY 7 1 ADC Advanced PC Link Software Installatio...

Страница 10: ...onnect your device and start with step 11 For more information on how to use HealthVault click Help at the top right hand corner of the page There you will find a detailed help document as well as an FAQ section to help with any issues you may have NOTE You cannot run the PC Link software and HealthVault at the same time Be sure to completely remove the PC Link software before installing HealthVau...

Страница 11: ...ther information Measurements should not occur immediately after each other otherwise the results will be skewed Wait for one minute in a relaxed position sitting or lying before you repeat a measurement 9 2 Delete Memory Before you delete all memory readings stored make sure you won t need to refer to them at a later date Keeping a permanent record is prudent and may provide additional informatio...

Страница 12: ...correctly seated Reposition the cuff and repeat the measurement ERR 5 HI LO The difference between systolic and diastolic is excessive Measure again carefully following proper procedures Consult your doctor if you still get unusual results of measurements The pressure in the cuff is too high over 300mmHg Relax for 5 minutes and repeat the measurement The pulse is too low less than 40 beats per min...

Страница 13: ... TECHNICAL SPECIFICATIONS Weight 35 lbs 148 g with batteries cuff Size 3 75 x 4 25 x 4 25 85mm x 77mm x 75mm including cuff Storage temperature 20 55 C 4 131 F Humidity 15 to 90 relative maximum humidity Operation temperature 50 F 104 F 10 C 40 C Display LCD Liquid Crystal Display Measuring method Oscillometric Pressure sensor Capacitive Measuring range Rated range of the determined Blood Pressure...

Страница 14: ...ulations of the International standard IEC60601 1 2 Clinical testing BHS A A Rated Clinical performance tests were carried out in the U K Association Protocol and British HTP Protocol 15 SYMBOL DEFINITIONS 16 HOW TO CONTACT US To register your product and obtain further detailed user information about our products and services visit us at www adctoday com and follow the links For questions comment...

Страница 15: ...ck Switzerland Read the instructions carefully before using this device Type BF applied part Dist by ADC 55 Commerce Drive Hauppauge NY 11788 Inspected in the USA Made in Taiwan tel 631 273 9600 toll free 1 800 232 2670 fax 631 273 9659 www adctoday com email info adctoday com IB p n 93 6016N 00 rev 2 english 0044 marking applies to model 6015N only ...

Отзывы: