When the remote display output is used with the Adam Equipment Remote Display unit, the
output is a continuous stream of data representing the weight and other information to
display the correct data on the remote display.
If the remote display data stream format is required for development purposes then please
contact the manufacturer for advice.
12.0 ERROR CHECKING
During weighing the balance is constantly checking to see if it is operating within the limited
parameters. The errors likely to occur are:
•
A/D counts below lowest allowed value
•
A/D counts above highest allowed value
•
A/D not operating
•
Maximum capacity exceeded
Other errors may be detected during special functions or operations. These will be described
in the section that applies.
Error messages and the reasons are:
Concerning A/D counts
ΕΡΡ ΥΛ
A/D counts below a limit
ΕΡΡ ΟΛ
A/D counts above a pre-set limit
Concerning calibration
ΕΡΡ ΣΤΒ
Calibration could not be completed
because the results were not stable
ΕΡΡ ΛΟ
or
ΕΡΡ ΗΙ
Calibration constant not within 20% of old calibration
constant
Concerning weighing
ΕΡΡ ΛΟ
Weight display is below zero by >4%max
ΕΡΡ ΗΙ
Weight is above maximum plus 90d