© Adam Equipment Company 2021
21
10.4.4
Animal (Dynamic) Weighing
The balance can be set to weigh animals or unstable/moving items. This is commonly
referred to as ‘Dynamic’ or ‘Animal’ weighing mode. The balance will collect the weight over
a period of time to arrive at an average value and display the value until the operator resets
the balance. The actual weighing process can begin either automatically when the weight is
placed on the pan, or when initiated by the operator. The weighing unit can be selected as
normal using the [Unit] and [Setup] keys, before starting the dynamic weighing process.
Steps:
Press [Mode] and then the [Up] or [Down] key to cycle through available modes. When the
“
Animal
” symbol is displayed, press [Setup] to enter animal weighing mode
. “
ρϖν
”
will now
be displayed on the screen.
Press [Up] or [Down] to select
“
ρϖν
”
for starting the dynamic weighing, or
“
ΣΕτϖΠ
”
to set
up the balance for dynamic weighing (see section 10.4.4.3 on Dynamic Weighing Setup
Parameters).
During dynamic weighing, if the [Print] key is pressed, the balance will display
“
ΠΑϖΣΕδ
”
for 1 second, then show the current average weight with the
“
Animal
”
symbol flashing
.
To resume, press [Print] again or if you do not wish to continue then pressing [Mode] will
display
“
ΣτΟΠ
”
for one second and then show the final value. The value will be locked until
the user releases it
.
10.4.4.1 Manual mode
When the balance is in the
“
ΜµΑΝϖΑΛ
”
mode:
–
If [Setup] is pressed when
“
ρϖΝ
”
is selected, balance will display
“
ΣτΑρτ
”.
Place the item on the pan and press [Setup] again.
After the pre-configured delay and test time have elapsed (see section 10.4.4.3 on Dynamic
Weighing Setup Parameters), the “
Hold
” symbol and the result will be displayed.
Remove the item from the pan. Press [Mode] to go back to
“
ρΥΝ
”
to weigh another item,
or [
0/T
] to return to normal weighing.
10.4.4.2 Auto mode
When the balance is in the
“
ΑϖΤΟ
”
mode:
–
If [Setup] is pressed when “
ρςΝ
is selected, the balance will display
“
ΛΟΑ∆ Παν
”.
Place the item on the pan. The animal weighing test will begin automatically.
After the pre-configured delay and test time have elapsed (see section 10.4.4.3 on Dynamic
Weighing Setup Parameters), the “
HOLD
” symbol and the result will be displayed.
Remove the item from the pan. Press [Mode] to go back to “
ρΥΝ
” to weigh another item, or
[
0/T
] to return to normal weighing.