5.0 ERROR CODES
During the initial power-on testing or during operation, the scale may show an error message. The meaning of the
error messages is described below.
If an error message is shown, repeat the step that caused the message. If the error message is still shown then
contact your dealer for support.
ERROR
CODE
DESCRIPTION
POSSIBLE CAUSES
Err 4
Err 4
Err 4
Err 4
Initial Zero is greater than allowed
(4% of maximum capacity) when
power is turned on or when [Zero]
is pressed.
Weight on the pan when turning the scale on.
Excessive weight on the pan when zeroing the scale.
Platform is not installed.
Improper calibration of the scale.
Damaged load cell.
Damaged Electronics.
Err 6
Err 6
Err 6
Err 6
A/D count is not correct when
turning the scale on.
Load cell is damaged.
Electronics is damaged.
Err 8
Err 8
Err 8
Err 8
High limit input error
Low limit is set first, then the high limit is set lower than
the low limit and high limit not equal to zero.
Err 9
Err 9
Err 9
Err 9
Low limit input error
High limit is set first, then the low limit is set higher than
the high limit and low limit not equal to zero.
FAIL H or
FAIL H or
FAIL H or
FAIL H or
FAIL L
FAIL L
FAIL L
FAIL L
Calibration error
Improper calibration (should be within +10% of the
factory calibration). The old calibration data will be
retained until the calibration process is complete.
Содержание WBW 1.5M
Страница 11: ...EN Adam Equipment Company 2014 9 3036610540RevB Nov14 7 0 MENU STRUCTURE ...
Страница 44: ...FR Adam Equipment Company 2014 42 3036610540RevB Nov14 18 0 ANNEXE ...
Страница 77: ...DE Adam Equipment Company 2014 75 3036610540RevB Nov14 18 0 ANHANG ...
Страница 89: ...ES Adam Equipment Company 2014 87 3036610540RevB Nov14 8 0 ESTRUCTURA DEL MENÚ ...
Страница 101: ...IT Adam Equipment Company 2014 99 3036610540RevB Nov14 7 0 STRUTTURA DEL MENU ...