12
700660229revA-Jul12
4) Press the
[Print]
key to change the setting for the printing detail control
and press
[Zero]
to save.
5) Press the
[kg/lb]
key to exit and continue.
6.5
ULTRASONIC
HEIGHT
CALIBRATION
FFFF5
5
5
5
UL
UL
UL
UL
TR
TR
TR
TR
A
A
A
A
The Ultrasonic height sensor has been factory calibrated and should not require
frequent calibration. If there is a problem with height measuring proceed as
below.
1) Switch on the scale then press the
[kg/lb]
key during the self-checking
test of the display.
2) The display will show the first function
F1
F1
F1
F1
EL.
EL.
EL.
EL.
3) Press the
[kg/lb]
key to select
FFFF5
5
5
5
U
U
U
U
LTRA
LTRA
LTRA
LTRA
and then
[Zero]
to enter the
sensor calibration section “AdJUST” XX (XX being a factory set figure)
4) Press the
[Zero]
key and the display will show “base- H 220”, press
[Zero]
to calibrate the “Zero” height reading.
5) The display will show “test” and a 4 digit number. Press the
[Zero]
key
from a position where the height sensor cannot detect you.
6) When calibration is complete the display will show “AdJUST XX”.
If after calibration the height sensor is not measuring correctly, press
[Print] or
[Clear]
at point 3 to change XX. A lower number will decrease the height
measurement and a higher number will increase it. Then continue from point 4
on.
*A print out with date and time can be issued following every calibration.
6.6
SETTING
THE
HOLD
FUNCTION
F6
F6
F6
F6
HOLD
HOLD
HOLD
HOLD
The Hold function can be set so that the height and BMI readings are visible as
they are being calculated
(OFF),
or remain as dashes until the weight reading is
stable upon where they will be displayed
(ON).
1) Switch on the scale then press the
[kg/lb]
key during the self-checking
test of the display.