© Adam Equipment Company 2006
7
forward to one position. When the figure displayed is same as the capacity of
the standard weight, press the
[On/Off]
key.
For example:
If a 10 kg weight is used instead of full capacity:
After you process till the step (9) as above, press the
[On/Off]
key.
The indicator will show
: “
0.0
”.
Press the
[Hold]
key, the last digit will be
automatically jumped from
“
0
”
to
”
9
”.
When the figure jumps to
“
1
”,
press the
[Hold]
key three times. Now, the
“
1
”
will be moved to the 3
rd
forward position
and the display will be changed to
“
10.0
”.
At this time you are required to
press the
[On/Off]
key immediately.
Next the display will show the current weight on the pan.
10. The Calibration is now complete.
11. Set the span jumper on the PCB to
OFF
position and re-seal the indicator.
12. If the selections made during calibration are not as described as above, the
scale may still work but will not show accurate results. In such cases, re-do
the calibration as shown above.