7.0 USER CALIBRATION
1. Press
[On/Off]
key when the display shows
”
SCALE
”.
2. Press
[On/Off]
key again.
If the display is not showing
“
Id01
”
open the upper housing and switch the
span jumper on PCB to
ON
.
Start again by pressing
[On/off]
.
The display shows
“
Id01
”, “
Id-02
”, “
Id05
”, “
Id10
”,”
Id-20
”, “
Id-50
”.
3. Select
“
Id01
”
for 100g increments and press the
[On/Off]
key once.
4. The indicator will show the selection capacity:
“
P1000
” “
P2000
”….”
P5000
”,
Select
“
P2500
”
for 250kg, then press the
[On/Off]
key once.
5. Indicator will automatically show the decimal points
: “
POT 0
” “
PT0.0
” “
T0.00
”
“
T0.000
”, “
POT 0
”.
Select the
“
PT0.0
”
and press the
[On/Off]
key again.
6. Indicator will display
“
ULOAD
”.
Press the
[On/Off]
key again.
7. Indicator will show the internal value. After it is stable, it will show
“
LOAd
”.
8. Put the mass (suggest the full capacity) and press the
[On/Off]
key, the
indicator shows the internal value (after loading) and the will show
“
Ld-In
”
after
stable.
9. Press the
[On/Off]
key. Then press the
[Hold]
key. The last digit of the
display will be moving from
“
0-9
”.
Press the
[Hold]
key, the digit will move