© Adam Equipment Company 2014
22
9520revF-Jun14
External Calibration
1)
Press the
[on/off]
key to turn the power on.
2)
Press the
[Smpl]
and
[Print]
keys at the same time during the self-checking
test.
3)
The display will show
unload.
Remove any weight from the stainless steel pan.
4)
When the stable indicator shows press the
[Tare]
key.
5)
The display will show the first weight that you can use to calibrate the unit.
You can change this value by pressing the
[Mode]
key. The weights that can
be used are as follows:
Model #
HCB123 HCB153
HCB302
HCB602 HCB602H HCB1002 HCB1502 HCB3001
Weight 1
60g
50g
100g
200g
200g
500g
500g
1000g
Weight 2
120g
100g
200g
400g
400g
1000g
1000g
2000g
Weight 3
-
150g
300g
600g
600g
-
1500g
3000g
6)
Once the calibration weight has been selected press the
[Tare]
key
7)
The display shows
LoAd.
Place your weight on the pan. Press
[Tare]
key
once
the stable sign is shown.
8)
The display will show
PASS.
Remove the weight from the pan.
NOTE:
If the calibration fails retry. The balance will show
FAllllL H
(when the weight is
higher) or
FAllllL L
(when the weight is lower). Repeat the process using the correct
calibration weight.
Содержание HIGHLAND HCB1002
Страница 2: ... Adam Equipment Company 2014 ENGLISH P1 30 FRANÇAIS P31 59 DEUTSCH P60 90 ESPAÑOL P91 116 ...
Страница 62: ... Adam Equipment Company 2014 60 9520revF Jun14 Adam Equipment HIGHLAND SERIE P N 9520 Revision E Oktober 2011 ...
Страница 89: ... Adam Equipment Company 2014 87 9520revF Jun14 10 0 KALIBRIER ZERTIFIKAT ...
Страница 93: ... Adam Equipment Company 2014 91 9520revF Jun14 Adam Equipment Serie Highland P N 9520 Revision D3 July 2010 ...
Страница 115: ... Adam Equipment Company 2014 113 9520revF Jun14 10 0 CERTIFICADO DE CALIBRACIÓN ...
Страница 119: ... Adam Equipment Company 2014 ...