© Adam Equipment Company 2014
17
9520revF-Jun14
8 n 1
8 data bits, no parity
7 E
1
7 data bits, even parity
7 0 1
7 data bits, odd parity
Press the
[Mode]
key to change the setting. Press
[Tare]
to confirm and move
to the next parameter.
9)
Format settings. The scale will print either using a format that includes
stability indications or one with only the weight. There are 2 settings that the
Highland can work with as follows:
FOr 1
Data format is with line headings and
extra line feeds as shown in section 9.3
FOr 2
Data format is weight only
Press the
[Mode]
key to change the setting. Press
[Tare]
to confirm and move
to the next parameter.
10)
The display will go back showing F3 SEr. Press the [Print] key to return to
weighing or press
[Mode]
to go to the next function.
6.4
AUTO
POWER
OFF
The auto power off function helps conserve power, when using the internal
rechargeable battery or AC adapter. The Auto switch-off time may be set up by the
user.
1)
To set this parameter press the
[Mode]
key during self-checking test when
turning on the balance.
2)
After a few seconds, the display will show the first function
F1 UNT
.
3)
Keep pressing the
[Mode]
key until
F4 oFF
is displayed.
4)
Press the
[Tare]
key to view the current settings.
5)
Press
[Mode]
to change the settings (0, 5, 10, 20 and 30 minutes).
6)
Press
[Tare]
to store a particular setting. The display will return to
F4 oFF
Содержание HIGHLAND HCB1002
Страница 2: ... Adam Equipment Company 2014 ENGLISH P1 30 FRANÇAIS P31 59 DEUTSCH P60 90 ESPAÑOL P91 116 ...
Страница 62: ... Adam Equipment Company 2014 60 9520revF Jun14 Adam Equipment HIGHLAND SERIE P N 9520 Revision E Oktober 2011 ...
Страница 89: ... Adam Equipment Company 2014 87 9520revF Jun14 10 0 KALIBRIER ZERTIFIKAT ...
Страница 93: ... Adam Equipment Company 2014 91 9520revF Jun14 Adam Equipment Serie Highland P N 9520 Revision D3 July 2010 ...
Страница 115: ... Adam Equipment Company 2014 113 9520revF Jun14 10 0 CERTIFICADO DE CALIBRACIÓN ...
Страница 119: ... Adam Equipment Company 2014 ...