Press the
[Mode]
key to change the setting. Press
[‐>0/T<‐]
to confirm and
move to the next parameter.
8)
Parity settings. Parity is a communications check. There are 3 settings that the
Highland can work with as follows:
8 n 1
8 data bits, no parity
7 E
1
7 data bits, even parity
7 0 1
7 data bits, odd parity
Press the
[Mode]
key to change the setting. Press
[‐>0/T<‐]
to confirm and
move to the next parameter.
9)
Format settings. The scale will print either using a format that includes
stability indications or one with only the weight. There are 2 settings that the
Highland can work with as follows:
FOr 1
Data format is with line headings and
extra line feeds as shown in section 9.3
FOr 2
Data format is weight only
Press the
[Mode]
key to change the setting. Press
[‐>0/T<‐]
to confirm and
move to the next parameter.
10)
The display will go back showing F3 SEr. Press the [Print] key to return to
weighing or press
[Mode]
to go to the next function.
6.4
AUTO
POWER
OFF
The auto power off function helps conserve power, when using the internal
rechargeable battery or DC adapter. The Auto switch‐off time may be set up by the
user.
1)
To set this parameter press the
[Mode]
key during self‐checking test when
turning on the balance.
2)
After a few seconds, the display will show the first function
F1 UNT
.