© Adam Equipment Company 2007
17
9.2 BACKLIGHT FOR LCD
•
The backlight of the LCD can be set to:
1. Be on all the time
2. Be on only when a weight is placed on the scale
3. Turned off
•
To set the backlight press and hold
[Pst]
key for 4 seconds.
•
The weight display will show “
EL xx
” where
xx
is the current
setting for the backlight.
•
Press
[U. Wt./Units]
to scroll through the options.
“
EL on
”
Sets the backlight to be on at all times.
“
EL Au
”
Sets the backlight to operate automatically when a weight is
placed on the scale or a key is pressed.
“
EL OFF
”
Sets the backlight to be off.
•
Press the
[Tare]
key to store the value or press the
[Zero]
key to escape from this setting and return to weighing.
10.0 ERROR CODES
During the initial power-on testing or during operation, the scale may
show an error message. The meaning of the error messages is
described below.
If an error message is shown, repeat the step that caused the message,
turning the balance on, carry out the calibration or other functions. If the
error message is still shown contact your dealer for further support.
ERROR CODE
DESCRIPTION
POSSIBLE CAUSES
Err 1
Time input error.
Tried to set an illegal time, i.e.
26hours
Err 2
Date input error
Tried to set an illegal date, i.e. 36
th
day
Err 4
Initial Zero is greater than
allowed (typically 4% of
the maximum capacity)
when power is turned on
or when the
[Zero]
key is
pressed,
Weight is on the pan when turning
the scale on.
Excessive weight on the pan when
zeroing the scale.
Improper calibration of the scale.
Damaged load cell.
Damaged Electronics.
Err 6
A/D count is not correct
when turning the scale
on.
Platform is not installed.
Damaged load-cell.
Damaged electronics.