© Adam Equipment Company 2007
16
•
During self-checking, press
[Zero]
and release at once. The
display shows “
SLEEP nodE
”.
•
Press
[U. Wt./Units]
key to scroll through the auto sleep
values.
“
0
” Auto sleep mode disabled
“
1
” Auto sleep after 1 minute
“
5
” Auto sleep after 5 minutes
“
10
” Auto sleep after 10 minutes
•
Press
[Tare]
to set the value. The scale returns to zero.
9.0 BATTERY AND BACKLIGHT OPERATION
9.1 RECHARGEABLE BATTERY
•
The scales can be operated from the battery, if desired. The
battery life is approximately 90 hours.
•
When the battery needs charging the arrow above the low
battery symbol under the “
Weight
” display will turn on. The
battery should be charged as soon as the arrow is on. The
scale will still operate for about 10 minutes after which it will
automatically switch off to protect the battery.
•
To charge the battery, simply plug the scale into the mains
and switch the mains power ON. The scale does not need to
be turned on.
•
The battery should be charged for at least 12 hours for full
capacity.
•
Just under the “
Count
” display is an LED to indicate the
status of battery charging. When the scale is plugged into
the mains power, the internal battery will be charged. If the
LED is green the battery is fully charged. If it is red, the
battery is nearly discharged and yellow indicates the battery
should be charged longer, preferably overnight.
•
If the battery has not been used properly or it is used for a
number of years it may eventually fail to hold a full charge. If
the battery life becomes unacceptable then contact your
distributor or Adam Equipment.