© Adam Equipment Company 2007
13
7.3 CALIBRATION
Next is the Calibration section. “
WEl
” will be displayed.
•
When
WEI
is displayed, press the
[Unit]
key to scroll through the
options for selecting the calibration mass.
•
To select the desired mass press the
[Mode]
key.
•
The raw ADC count will be displayed along with “
CAI
”.
•
To set the zero calibration, press the
[Mode]
key again.
•
The scale will ask you to load the calibration mass by displaying the
new ADC and “
LOD
”
•
Place the calibration mass (same as the chosen “
WEI
” capacity) and
when the stable indicator is on, press
[Mode]
again.
•
The display will show the value of the calibration mass.
•
Remove the weight.
•
The scale will return to the normal weighing.
Содержание BFW 150
Страница 2: ...Adam Equipment Company 2007 2...