35
Acute Technology Inc.
Copyright
2018
1.
Schematic diagram of the current set of trigger conditions
2.
Trigger conditions setting
As shown in the following diagram, the first and the second classes are continuous
trigger, the relationship between the second and third classes are non-continuous
trigger, and the third, fourth, fifth and sixth classes are continuous trigger.
Difference between the continuous trigger and the non-continuous trigger lies in:
Continuous trigger: The signal captured by two adjacent sample clocks must meet the
conditions to trigger.
Non-continuous trigger: It is triggered only when both the first condition and the
second condition are met, no matter how many signals appear
in between the first condition and the second condition.
Therefore, such a trigger condition is not continuous at all.
A continuous trigger mode is usually set when Synchronous or State is used for
measurement, because the use of synchronous clock is usually in a measurement state,
and the signal is in a continuous state. Under the Asynchronous or Timing situation, it
is common for the signal at the changing edge to meet the continuous triggering
condition, while it is difficult for most of the signals to meet the conditions of
continuous triggering, and therefore it is suitable to set non-continuous trigger as a
condition for them.