For service assistance, please call your authorized ActiveCare dealer.
19
Prowler 3310/3410 Owner’s Manual
Table 2: Scooter Controller Internal Diagnostics
Flash
code
Possible Cause
Solution
1
The battery needs charging.
The battery voltage has dropped below
23.3 volts in neutral. Recharge
batteries.
2
Battery voltage is too low.
The battery has dropped past 16.5
volts and is not sufficient to allow safe
driving. Check the battery condition
and the connections.
3
Battery voltage too high
The battery has exceeded 32 volts.
Check the battery condition and the
connections.
4
Current limit time out
The motor current has reached too
high a value. Check the condition of
the motor and loom. Contact your
Service Agent.
5
Park brake fault
Check park brake conditions.
6
Speed control out of neutral
Return speed pod to neutral and rest
system. Readjust the speed lever to
neutral if necessary. Contact your
Service Agent.
7
Speed control fault
Check speed pod wiring for open or
short circuits. Check speed pot set-up.
Contact your Service Agent.
8
Motor fault
Contact your Service Agent.
9
Internal (controller) fault
Contact your Service Agent.