background image

BEFORE YOU START

5

4

The Choice of Power Supply

1.Battery powered mode:

    6VDC 4*AAA alkaline batteries

Installing and Replacing the Batteries

In order to achieve the best performance and protect your monitor,

please use the authorized / specified battery

CAUTION

1. Open the battery door.

2. Insert the batteries according to the polarity 

    indications. 

3. Close the battery door. 

Battery Life: Approx. 44 days

(Battery capacity: 600 mAH. If measured three times per day, each measurement takes 35s, 

measuring result display takes 20s and data transmission takes 10s. The current for measurement is 

400 mA and that for records display and data transmission is 50 mA and 50 mA separately, while the 

current when shutdown is 35 uA)

Replace the batteries under following circumstances:

displays on the LCD.

The LCD display dims.

When powering on the monitor, the LCD doesn’t light up.

Remove batteries if the device is not likely to be used for some time.
Worn batteries are harmful to the environment. Do not dispose with daily garbage.
Dispose old batteries following your local recycling guidelines.

CAUTION

BEFORE YOU START

Setting Date, Time and Measurement Unit

1.When the monitor is OFF, press 

and hold “User 1” button to enter 

Time Setting Mode.

Please proceed to time setting before your initial use so as to ensure 

each record is labeled with a time stamp. 

NOTE: The monitor will shut off in 60 seconds after last 

operation when in Setting Interface.

2.As pictured on the right, the 

blinking numeral represents the 

[HOUR]. Press “Query” button to 

change the numeral. Each press 

will increase the numeral by one 

in a cycling manner.

3.Press “User 1” button to confirm 

the [HOUR]. Then the monitor 

diverts to [MINUTE] setting . 

 MINUTE 

 HOUR

Содержание ACMMD002

Страница 1: ...oughly so as to ensure the safe usage of this product and keep the manual for further reference in case you have problems Thank you very much for selecting the Actiiv Blood Pressure Monitor ACMMD002 MODEL ACMMD002 BLOOD PRESSURE MONITOR User Manual ...

Страница 2: ...lease read this user manual carefully and thoroughly before use This device is intended for adult use in the home domestic environment only This device is intended for non invasive measuring and monitoring of arterial blood pressure It is not intended for use on extremities other than the upper arm or for functions other than obtaining a blood pressure measurement Do not confuse self monitoring wi...

Страница 3: ...rtbeat detection during the measurement Arrhythmia Beats per minute BPM Pulse per minute 1mmHg 0 133kPa 1kPa 7 5mmHg SYMBOL DESCRIPTION EXPLANATION Systolic Blood Pressure High blood pressure Diastolic Blood Pressure Low blood pressure User 1 User 2 Data Pending to Transmit Measurement data stored in the equipment Shaking reminder Shaking will result in inaccurate measurements Low Battery Low batt...

Страница 4: ...g circumstances displays on the LCD The LCD display dims When powering on the monitor the LCD doesn t light up Remove batteries if the device is not likely to be used for some time Worn batteries are harmful to the environment Do not dispose with daily garbage Dispose old batteries following your local recycling guidelines CAUTION BEFORE YOU START Setting Date Time and Measurement Unit 1 When the ...

Страница 5: ...oid Phone Simple and convenient With the advanced Bluetooth 4 0 technology applied to this blood pressure monitor mobile phones which are equipped with Bluetooth functionality are able to receive your personal health data Just simply install the specially designed app and pair up your blood pressure monitor with your mobile phone by following the instructions below 2 Turn on Bluetooth and the App ...

Страница 6: ...e of the cuff should remain at the same level as your heart Your legs should be relaxed with feet falling outwards Palms up Start Measurement Start the Measurement 1 LCD display Tips 3 Inflating and measuring 5 2 Adjust to zero Press User 1 User 2 button to turn off the monitor Otherwise the monitor will shut off within 1 minute after last operation 4 Display and save the results and the data will ...

Страница 7: ...s Query button to confirm deletion from memory The LCD will display dEL dOnE indicates that the deletion is complete and then turn off Data Transmission 1 With the device successfully paired up with your iPhone the measurement data will be automatically transmitted to your mobile via Bluetooth 2 The symbol will disappear after successful data transmission and you may check your personal health data...

Страница 8: ...ry cold environemnt Immediate measurement after tea coffee smoking When talking or moving your fingers When you want to discharge urine Disposal Degraded sensors may result in inaccurate measurement while loosened electrodes may cause the monitor s failure to power on The expected life of the monitor is two years Please dispose of accessories detachable parts and the ME equipement according to the ...

Страница 9: ...ing Diastolic What is the standard blood pressure classification When ventricles contract and pump blood out of the heart the blood pressure reaches its maximum value in the cycle which is called systolic pressure When the ventricles relax the blood pressure reaches its minimum value in the cycle which is called diastolic pressure Irregular Heartbeat Detector This Blood Pressure Monitor is equipped...

Страница 10: ...nd frequently asked questions for problems you may encounter with your blood pressure monitor If the product is not operating as you think it should check here before arranging for servicing PROBLEM SYMPTOM CHECK THIS REMEDY No power Low batteries Error message Display will not light up Batteries are exhausted Replace with new batteries Insert the batteries correctly Replace with new batteries Bat...

Страница 11: ...e Table 1 Guidance and MANUFACTURER S declaration ELECTROMAGNETIC EMISSIONS for all ME EQUIPMENT and ME SYSTEMS Guidance and manufacturer s declaration electromagnetic emissions Emissions test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 2 RF emissions CISPR 11 Class B Harmonic emissions IEC 61000 3 2 Not applicable Not applicable Voltage fluctuations flicker emissions...

Страница 12: ...RER S declaration electromagnetic IMMUNITY for all ME EQUIPMENT and ME SYSTEMS Table 4 Guidance and MANUFACTURER S declaration electromagnetic IMMUNITY for ME EQUIPMENT and ME SYSTEMS that are not LIFE SUPPORTING EMC GUIDANCE EMC GUIDANCE ...

Страница 13: ...u have signed in 6 Press Edit at the bottom right corner untick Weight press the back button and Yes to save Data this will remove the weight measurement and return to the Start menu 7 Press the Settings button in the top right corner 8 Press the Add button to add a device 9 Press the Bluetooth logo 10 Please ensure you pressand hold the Start button on the blood pressure monitor to initiate Pairi...

Отзывы: