![Acteon SOPRO 717 FIRST Скачать руководство пользователя страница 9](http://html1.mh-extra.com/html/acteon/sopro-717-first/sopro-717-first_user-manual_2842333009.webp)
ENGLISH
SOPRO 717 First • User’s manual
8
4.5. ELECTROMAGNETIC COMPTABILITY
Guide and declaration of the manufacturer - electromagnetic emissions
SOPRO device is intended to be used in the electromagnetic environment speci
fi
ed below.
The user should make sure it is used in this environment.
Emission trial
Compliance
Electromagnetic environment - Guide
RF emissions
CISPR 11
Group 1
SOPRO device only uses radio energy for its internal functions.
Therefore, its RF emissions are very low and are unlikely to cause
interference with nearby electronic devices.
RF emissions
CISPR 11
Class B
SOPRO device may be used in all domestic environments, including
the ones directly connected to the public low voltage power
distribution network used to supply household buildings.
Harmonic emissions
EN 61000-3-2
Not applicable
Voltage
fl
uctuations / Flicker
EN 61000-3-3
Applicable
Содержание SOPRO 717 FIRST
Страница 1: ...User s manual F I R S T...