![Acteon SOPRO 717 FIRST Скачать руководство пользователя страница 10](http://html1.mh-extra.com/html/acteon/sopro-717-first/sopro-717-first_user-manual_2842333010.webp)
ENGLISH
SOPRO 717 First • User’s manual
9
Guide and declaration of the manufacturer - electromagnetic immunity
SOPRO device is intended to be used in the electromagnetic environment speci
fi
ed below.
The user should make sure it is used in this environment.
Immunity trial
CEI 60601
Severity level
Compliance
level
Electromagnetic environment
Guide
Electrostatic
discharges
EN 61000-4-2
± 6 kV when in contact
± 8 kV in the air
± 6 kV
± 8 kV
The
fl
oor should be wooden, concrete or tile. If the
fl
oor is covered with a synthetic material, the relative
humidity should be at least 30%.
Far transient bursts
EN 61000-4-4
± 2 kV for the feed cables
± 1 kV for the input/out-
put cables
± 2 kV
± 1 kV
The main power supply quality should be one of a
traditional commercial or hospital environment.
Voltage shocks
EN 61000-4-5
Differential mode ± 1 kV
Common mode ± 2 kV
± 1 kV
N.A.
The main power supply quality should be one of a
traditional commercial or hospital environment.
Dips, brief outages
and power voltage
variation
EN 61000-4-11
• <5% Ut - for 10 ms
• 40% Ut - for 100 ms
• 70% Ut - for 500 ms
• <5% Ut - for 5 s
<5% Ut 10 ms
<40% Ut 100 ms
<70% Ut 500 ms
<5% Ut 5 s
The main power supply quality should be one of a
traditional commercial or hospital environment.
If the user of SOPRO device requires it to continue
to operate during main power supply outages, it is
recommended SOPRO device is fed by an inverter or a
battery.
Magnetic
fi
eld with the
network frequency
(50/60 Hz)
3 A/m
3 A/m
The magnetic
fi
eld with the network frequency should
be at a characteristic level of a location in a traditional
commercial or hospital environment.
Note:
Ut is the power voltage nominal value applied during the trial.