8 Disinfection and sterilising
The instructions relating to cleaning, disinfection and sterilisation protocols for accessories provided by SATELEC, a
company of Acteon group have been approved for each medical device and accessory. The applicable guides are listed
in the chapter
Associated documentation page 5
.
They can be downloaded at the following address:
www.satelec.com/documents
.
In all cases, the local regulations in force relating to the cleaning, disinfection and sterilisation protocols for
accessories take precedence over the information provided by SATELEC, a company of Acteon group.
8.1 Clean and disinfect the medical device
The medical device must be in OFF or O stop position during cleaning and disinfecting procedures.
Avoid using cleaning and disinfection products that contain flammable agents.
Otherwise, ensure that the product has completely evaporated or that there is no fuel left on the medical device and its
accessories before switching it on.
Do not use an abrasive product to clean the medical device.
Never apply sprays directly to the medical device to clean it. Always spray the product onto a wipe, then clean
the medical device.
The medical device control unit, handpiece cord and control pedal must be cleaned and disinfected daily.
Use alcohol disinfectant wipes.
8.1.1 Pre-disinfection and cleaning – Manual method
Minimum duration of step
Cleaning instructions
Clean with alcoholic disinfecting wipes.
Inspect visually.
Repeat the preceding steps until there is no more visible contamination.
at least 2 minutes
Rinse in demineralised water.
Dry using a soft lint-free cloth or medical grade clean compressed air
Inspect visually.
Repeat the preceding steps until there is no more visible contamination.
Always empty the powder container if the device will not be used for several hours (e.g., at night). The residual humidity
diminishes the properties of the powder and increases the risk of clogging.
Proceed as follows:
1. Switch off the medical device.
2. Remove the container from the top.
3. Collect the powder with the suction cannula.
4. Clean the container and the plug.
5. Close the tank.
8.2 Cleaning the irrigation system
It is imperative to purge the Prophypen handpiece and cord after every use, no matter how short.
Press and hold the Purge
button to activate the purge.
The air circuit is active for a few seconds, then stops automatically.
User Manual | AIR MAX | J11401 | V7 | (03) | 12/2017 | ND07EN010G - Page 21/42
Содержание Air Max
Страница 1: ...User Manual AIR MAX ...
Страница 6: ...Page 6 42 User Manual AIR MAX J11401 V7 03 12 2017 ND07EN010G ...
Страница 10: ...Page 10 42 User Manual AIR MAX J11401 V7 03 12 2017 ND07EN010G ...
Страница 12: ...Page 12 42 User Manual AIR MAX J11401 V7 03 12 2017 ND07EN010G ...
Страница 18: ...Page 18 42 User Manual AIR MAX J11401 V7 03 12 2017 ND07EN010G ...
Страница 20: ...Page 20 42 User Manual AIR MAX J11401 V7 03 12 2017 ND07EN010G ...
Страница 26: ...Page 26 42 User Manual AIR MAX J11401 V7 03 12 2017 ND07EN010G ...
Страница 30: ...Page 30 42 User Manual AIR MAX J11401 V7 03 12 2017 ND07EN010G ...
Страница 38: ...Page 38 42 User Manual AIR MAX J11401 V7 03 12 2017 ND07EN010G ...