7 Maintenance
The only preventive maintenance the medical device requires is:
l
monitoring of accessories;
l
everyday cleaning, disinfection and sterilisation;
l
cleaning.
7.1 Checking the lamp power level
It is important to regularly check that the lamp is working correctly.
This can be done using a purpose-designed tester.
To do this, proceed as follows:
l
Check that the optical guide is intact and has no composite residues.
l
Set the lamp to Rapid mode.
l
Insert the optical guide into the power tester.
l
Switch on the Mini LED.
l
The power tester may give the following results:
Colour
Result
Green The lamp is working correctly and has a power output equal to or more than 785 mW per cm
2
Red
The lamp is not working correctly. Please read the chapitre
Identifying incorrect operation page 19
7.2 Identifying incorrect operation
In the event of incorrect operation, refer to the tables below to quickly identify and repair the non-complex parts of the
medical device.
If the incorrect operation is not described in the tables below, please contact your supplier or the customer service
team at SATELEC, a company of Acteon group.
Do not use the medical device if it appears to be damaged or faulty. Isolate the medical device and make sure that it
cannot be used.
7.2.1 Not working
Symptoms: The Mini LED does not switch on and does not emit blue light.
Possible causes
Solutions
The on/off button is set to off
Press the on/off button to switch on the Mini LED
The battery is flat
Charge the battery
The charging base is incorrectly connected to the mains
power preventing the battery from charging normally
l
Disconnect the charging base;
l
Check that the socket is not faulty;
l
Connect the charging base;
l
Check that the charging light comes on;
l
Wait for the battery to fully charge.
The battery switched itself to safety mode
Install another battery
The internal temperature of the Mini LED has reached the
maximum permitted level
Allow the Mini LED to cool before using it again
The battery charging light is red. The battery was not fully
charged before being switched on
Remove the partly charged battery and install a fully
charged battery
7.3 The power of the Mini LED is not as expected
Possible causes
Solutions
Cure composite residue remains on the optical guide
l
Remove the residue.
l
Check that the surface of the optical guide is intact.
l
Change the optical guide if necessary.
User Manual • Mini LED™ • J02541 • V6 • (07) • 07/2016 • NG13EN010F - Page 19/30
Содержание Mini LED
Страница 1: ...User Manual Mini LED ...
Страница 6: ......
Страница 8: ...Page 6 30 User Manual Mini LED J02541 V6 07 07 2016 NG13EN010F ...
Страница 10: ...Page 8 30 User Manual Mini LED J02541 V6 07 07 2016 NG13EN010F ...
Страница 18: ...Page 16 30 User Manual Mini LED J02541 V6 07 07 2016 NG13EN010F ...
Страница 20: ...Page 18 30 User Manual Mini LED J02541 V6 07 07 2016 NG13EN010F ...
Страница 30: ...Page 28 30 User Manual Mini LED J02541 V6 07 07 2016 NG13EN010F ...