Waveform (nominal value, irradiance, duration)
Average irradiance over time
Pulsed mode
1020 mW / cm
2
754 mW / cm
2
Progressive mode
937 mW / cm
2
693 mW / cm
2
4.9.3 Cure depth
Cure depth in mm with a distance of 2 mm between the optical guide and the cure material
Rapid mode
Pulsed mode
Progressive mode
Z100 MP, 3M
3 mm/2.5 mm
2 mm/2.5 mm
3 mm/2.5 mm
Tetric EvoCeram, Ivoclar Vivadent
2 mm/1.5 mm
1.5 mm/1.5 mm
2 mm/2 mm
Charisma, Heraeus - Kulzer
2 mm/2 mm
2 mm/1.5 mm
2.5 mm/2 mm
4.10 Switching off the medical device
Press the ON/OFF button to stop a cure cycle.
The Mini LED switches off after three minutes of inactivity. Press the button to restart.
4.11 Disconnecting the medical device
Before a long absence or when not in use, the medical device must be cleaned, its battery must be removed and the
charging base must be disconnected from the mains power.
Page 14/30 - User Manual • Mini LED™ • J02541 • V6 • (07) • 07/2016 • NG13EN010F
Содержание Mini LED
Страница 1: ...User Manual Mini LED ...
Страница 6: ......
Страница 8: ...Page 6 30 User Manual Mini LED J02541 V6 07 07 2016 NG13EN010F ...
Страница 10: ...Page 8 30 User Manual Mini LED J02541 V6 07 07 2016 NG13EN010F ...
Страница 18: ...Page 16 30 User Manual Mini LED J02541 V6 07 07 2016 NG13EN010F ...
Страница 20: ...Page 18 30 User Manual Mini LED J02541 V6 07 07 2016 NG13EN010F ...
Страница 30: ...Page 28 30 User Manual Mini LED J02541 V6 07 07 2016 NG13EN010F ...