А1270 Electro-Magnetic Ultrasonic Thickness Gauge
Operation Manual
30
ACOUSTIC
CONTROL
SYSTEMS
To calibrate the velocity:
- Set the thickness of the sample.
- Start the procedure of velocity calibration with the sample.
- The message "Put the probe on calibrating sample and press ENTER"is displayed on the screen.
- Apply some coupling liquid on the sample.
- Place EMAT on the sample and press
.
- The message "Data acquisition on the sample thickness of XX.XX mm", where ХХ.XХ is the set thickness of the sample, is
displayed on the screen.
Figure 28
2
Содержание A1270 EMAT