background image

 

 

 

SAT 

 

page 3 of 9 

 

 

Application field 

 

The electrode probe 

SAT

, in combination with a suitable evaluation device, is used for conductive limit value 

detection in electrically conductive filling materials. 

 

Depending on the application, the electrode rods can have a length of up to 3 m. 

 

With up to four contact electrodes multiple tasks can be fulfilled simultaneously. Among these tasks are e.g. 
leakage and overflow protection, minimum / maximum protection resp. multilevel detection in container, pump 
protection in pipelines and also the two-position-control of pumps. 

 

The electrode probes are designed for a wide application range. 
The conductivity also of aggressive filling materials, with 1 µS/cm and higher can be detected, at process 
temperatures from -40 °C to +150 °C, at pressures from -1 bar to +20 bar. 

 

The electrode probes are certificated for the use in explosion hazardous areas acc. to ATEX II 1 G in zone 0. 

 
 
 
 

 
 
 
 
 
 
 
 

 
 
 
 
 
 

Function 

 

The electrode probe is mounted directly in the wall of the container or of the pipe by using the respective process 
connection or installed over the filling material by using a suitable mount. 

 

The alternating voltage, that is generated by a suitable evaluation electronic is applied either between the 
electrode rods or between the electrode rods and the metallic wall of the container resp. pipe that is connected 
to the metallic process connection. 

 

As soon as the electrically conductive filling material makes a connection between the electrodes resp. between 
the electrode and the metallic wall of the container resp. pipe, an measurable current flows, that causes a 
reaction of the connected evaluation device. 

 

Due to the use of a alternating voltage the corrosion at the electrode and the electrolytic decomposition of the 
filling material is avoided. 

 

At the electrode probes an additional module (diode module LBM) for the wire supervision can be installed inside 
the housing. 
In the case of a wire break between electrode probe and a suitable evaluation device, the evaluation device can 
output a corresponding warning signal. 

 
 

 
 
 

Содержание SAT

Страница 1: ...position control of pumps Wide application range for conductivities higher than 1 S cm for process temperatures from 15 C to 150 C for process pressures from 1 bar to 10 bar materials also for aggres...

Страница 2: ...SAT page 2 of 9 Index Application field 3 Function 3 Safety notes 3 Installation 4 Maintenance 4 Repair 5 Electrical connection 5...

Страница 3: ...trode probe is mounted directly in the wall of the container or of the pipe by using the respective process connection or installed over the filling material by using a suitable mount The alternating...

Страница 4: ...ce is installed and operated in explosive hazardous areas the general Ex construction standards EN IEC 60079 14 VDE 0165 these safety notes and the enclosed EC conformity certificate must be observed...

Страница 5: ...y if the filling material is strongly fluctuating Use for the stabilization suitable isolating spacers The distance between the spaces should be not more than 0 5 m At horizontal side mounting into a...

Страница 6: ...parated from power leading wires At strong electromagnetic irradiation use principally a cable with shield Connect the cable shield only at one side to earth A suitable evaluation device must by conne...

Страница 7: ...5 C Limitation by material Environmental temperature range Terminal enclosure PP 5 100 C Process temperature maximum 15 C 150 C observe limitations Limitation by variant Process temperature range ATEX...

Страница 8: ...rocess connection G15 G 1 Process connection G20 G 2 Process connection G10 G 1 Process connection G12 G Terminal enclosure 60mm only with material POM Terminal enclosure 40mm with terminal enclosure...

Страница 9: ...others on request Material process connection terminal enclosure D POM 40mm for G G 1 resp 80mm for G 1 G 2 E POM 60mm for G G 1 P PP 40mm for process connection G G 1 M PP 80mm for process connection...

Отзывы: