18
8 Typical applications
The application of medical residual current monitoring products in the operating room power distribution system
is shown in the figure below:
Note: The grounding in the isolated power supply cabinet should be connected reliably with the equipotential
terminals in the field.
9 Power on and debugging instructions
9.1 Wiring check
Each set of residual current monitoring products should be connected before power on, mainly to check
whether there is wrong connection, missing connection or short circuit. According to the wiring diagram shown in
part 5.4 of this manual, the inspection can be carried out in the following order:
(1) Check each set of residual current monitoring system, including AIM-R100 residual current monitor,
AKH0.66/l-20 residual current transformer, AID150 centralized alarm and display instrument, and ACLP10-24
DC24V power supply which supply power for AID150.
(2) Check input terminals 1 and 2 of ACLP10-24 DC24V power supply in each set of residual current
products are connected to AC220 V power supply. Whether the 24V and G of the output end are reliably
connected with the 24 V and G terminals of the AID150 centralized display instrument, and the positive and
negative poles are correct.
C65N C16/2P
C65N-C16/2P+Vigi C65 ELM 32A/2P 30mA
ATSE
BG3-63II/N4DR
Interpact
INS 63A
PZ300
AKH-0.66Z
80/0.05
ACLP10-24
AID150
DC 24V
RS485
C65H ICB 16A/2P
C65H ICB 16A/2P
C65H ICB 16A/2P
C65H ICB 20A/2P
C65H ICB 16A/2P
C65H ICB 16A/2P
C65H ICB 16A/2P
C65N-C16/2P+Vigi C65 ELM 32A/2P 30mA
C65N-C16/2P+Vigi C65 ELM 32A/2P 30mA
C65N-C16/2P+Vigi C65 ELM 32A/2P 30mA
C65N-C16/2P+Vigi C65 ELM 32A/2P 30mA
C65N-C16/2P+Vigi C65 ELM 32A/2P 30mA
C65N-C16/2P+Vigi C65 ELM 32A/2P 30mA
C65N-C20/4P+Vigi C65 ELM 32A/4P 30mA
6A
WX1
WX2
WX3
WX4
WX5
WX6
WX7
WX8
C65 ICB
40A/2P
WL1
WL2
WL3
WL4
WL5
WL6
WL7
WL8
SPD
WL9
AC220V/AC220V 8kVA
Operating room isolation power cabinet
Medical isolation transformer
L
o
a
d
I
n
s
u
l
T
e
m
p
Alarm and display instrument
(Installed in operating room)
Wall socket1 (special for surgery)
Wall socket2 (special for surgery)
Wall socket3 (special for surgery)
Wall socket4 (special for surgery)
Tower crane
Central Intelligence Control Panel
Operating Bed
PC Platform
(AC220/24V)
Emergency lamp
Viewing lamp
Writing table lamp
Operation table
OR1 lighting
Three-phase socket ( reserve)
OR2 lighting
Electric door power supply
Operating room shadowless lamp
Special power supply
AIM-R100
AKH-0.66/L-20
Residual Current Monitor
C65 ICB 40A/2P
AITR8000
AIM-M100
C65H ICB 16A/2P
AKH-0.66P26