8
SYNCHRONY EQUIPMENT USER MANUAL
© COPYRIGHT 2015-2021, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
CONFIDENTIAL AND PROPRIETARY
Pre-exercise process – Resting baseline
All initial exercises will start with a pre-exercise session unless the clinician indicates differently. The software
prompts the patient to remain still and quiet for 5 seconds to obtain the “resting baseline.” To capture the resting
baseline manually, select “Mark manually”. To skip the resting baseline, select “Skip baseline”
Pre-exercise process – Mark 3 repetitions
The “Mark Repetitions” step will calibrate the exercise specific to the patient’s needs. Capturing three (3) typical
swallows prior to beginning of the effortful swallow exercise customizes the exercise target level for the specific patient.
Automatic
–
requests patient to
remain still and quiet for 5 seconds
Mark manually
– the therapist
controls the start and stop time of the
resting baseline
Skip baseline
– the resting baseline
will default to 10mV
Note –
when footplate is attached the
footplate command L-M-R are
displayed on the command boxes.
Mark repetition –
use the footplate or
mouse to mark 3 typical swallow
repetitions.
Remove last –
allows the therapist to
remove the last captured/marked repetition
Pause/menu –
pauses the pre-exercise
process
Содержание Synchrony
Страница 4: ......