OMNISTAND™ USER MANUAL
15
COPYRIGHT © 2016, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED.
Installing the tread extension
Secure tread extension plate
M
into screws at double position
L
.
Note:
Moving the OmniStand™ with the tread extension plate installed is not possible. Move the
OmniStand™ only when the tread extension is removed.
Setting dynamic standing function
Before every training session:
Lift OmniStand™ by upper pipe frame on left and right side.
If it is possible to remove the vertical pipe from the balance unit, then training with the OmniStand™ is not
allowed. Contact ACP Customer Service.
Set release lever
N
to 2, 1 or 0 on both vertical pipes. Make sure that release lever latches in.
Assure release lever is set to the same position on both sides
Position 2
= dynamic standing function
enabled with full range of
movement (up to 11°)
Position 1
= dynamic standing function
enabled with limited range of
movement (up to 6°)
Position 0
= dynamic standing function
disabled
N