6
3.2
Safety Instructions
Please comply with the safety instructions compiled below to prevent fire, electrical shocks, injuries and damage to the
unit.
Protect the unit from exposure to strong infrared radiation like i.e. direct impact of sunlight.
The unit may only be operated under the following ambient conditions:
Store the unit for three hours at the new operating location before commissioning if a difference in temperature of 10°C is
exceeded. The unit might be damaged due to fast changes in temperature (i.e. by condensed water).
Prevent the unit from showing the same screen content over a longer period of time because TFT displays might have
after images due to the image-sticking effect. Use the Display Off function key to turn the display off if you do not use
the unit for a longer period of time (this does not have any impact on the running programs or functions of the unit). A
LED in the operating panel indicates the fact that the display is deactivated. Alternatively a screensaver could be used.
Protect the unit from strong magnetic fields like the field close to a MRI.
Protect the unit from exposure to strong electromagnetic radiation. Operation close to HF surgery equipment is allowed
for units with optional cable cover as long as the minimum distances listed in Abschnitt 3.3 are maintained.
Protect the unit from exposure to strong ionizing radiation like i.e. the radiation at a C-arm on X-ray devices.
Furthermore, the unit shall be protected from strong impacts and ongoing vibrations.
Temperature
Storage
Humidity
Operating
0 to 40°C
*1
-10 to 60°C
10 to 90%
non-condensing
Altitude
5 to 70%
non-condensing
0 to 3.048m
0 to 10.000ft
0 to 12.192m
0 to 40.000ft
Pressure
105 to 70kPa
105 to 19kPa