10.11 During the operations performed in the region of the thorax or the head it is
recommended to avoid using the flammable anaesthetics or oxidising gases such as
nitrous oxide (
N
2
O)
and oxygen, unless those agents are aspirated away.
10.12 In order to remove gases and increase the visibility during the operation, use of
smoke-plume extraction is recommended, where removal is not possible in any other
way.
10.13 Use non-flammable disinfectants. Otherwise, they should be left to evaporate
before the procedure. There is also a risk of pouring those agents under the body or into
a body cavity. Should this happen, such flooded areas should be dried. A flammable
agent can be set on fire by a spark occurring during normal system operation.
10.14 Sparks at the active electrode present the risk of setting bandages and metabolic
gases on fire.
10.15 During a procedure, there is also a risk that a heart pacemaker can be damaged or
its function can be interfered during a procedure. In these cases, the bipolar technique
should be used. If monopolar modes are necessary, the neutral electrode should be
placed at a possibly greatest distance from the pacemaker. The active electrode should
not be used near the stimulator. It is recommended to apply the current for a short time
at short time intervals. Before using electrosurgery, consult an authorised representative
for the heart pacemaker and a cardiac surgeon. Check the pacemaker thoroughly after
the procedure. Use of electrosurgery systems on patients with heart pacemakers is not
permitted under outpatient clinic conditions.
10.16 High-frequency leakage currents can cause burns at a distance from the electrode
application site, if they are in contact with conductive elements.
10.17 The commonly-used “through-the-instrument” coagulation technique should only
be applied when using properly insulated forceps. These are special forceps with
insulated handles.
Surgical gloves do not sufficiently protect the operator from
burns.
Never use spray coagulation when applying this technique.
10.18 When using spray coagulation, keep at an appropriate distance from the fingers,
metal parts of endoscope optics, fiberscopes.
10.19 When performing endoscopic procedures:
IMPORTANT NOTICE
maintain the active part of the electrode in the operator’s field of vision to avoid
accidental burns or coagulation at a random site
avoid contact with the metal parts of the endoscope
use a non-conductive cap on the endoscope eyepiece
10.20 In designing electrosurgical generators, Ackermann paid special attention to the
increasingly restrictive requirements regarding electromagnetic emissions. As a result,
solutions that ensure minimal emission levels were selected to fulfil current and future
requirements.
Page: 94 of 134
Содержание 16-2000-700
Страница 99: ...USER MANUAL 16 2000 700 701 702 703 Rev 05 2018 Page 99 of 134 ...
Страница 100: ...USER MANUAL 16 2000 700 701 702 703 Rev 05 2018 Page 100 of 134 ...
Страница 105: ...USER MANUAL 16 2000 700 701 702 703 Rev 05 2018 Page 105 of 134 ...
Страница 106: ...USER MANUAL 16 2000 700 701 702 703 Rev 05 2018 Page 106 of 134 ...
Страница 107: ...USER MANUAL 16 2000 700 701 702 703 Rev 05 2018 Page 107 of 134 ...
Страница 108: ...USER MANUAL 16 2000 700 701 702 703 Rev 05 2018 Page 108 of 134 ...
Страница 109: ...USER MANUAL 16 2000 700 701 702 703 Rev 05 2018 Page 109 of 134 ...
Страница 110: ...USER MANUAL 16 2000 700 701 702 703 Rev 05 2018 Page 110 of 134 ...
Страница 111: ...USER MANUAL 16 2000 700 701 702 703 Rev 05 2018 Page 111 of 134 ...
Страница 112: ...USER MANUAL 16 2000 700 701 702 703 Rev 05 2018 Page 112 of 134 ...
Страница 113: ...USER MANUAL 16 2000 700 701 702 703 Rev 05 2018 Page 113 of 134 ...
Страница 114: ...USER MANUAL 16 2000 700 701 702 703 Rev 05 2018 Page 114 of 134 ...
Страница 115: ...USER MANUAL 16 2000 700 701 702 703 Rev 05 2018 Page 115 of 134 ...
Страница 116: ...USER MANUAL 16 2000 700 701 702 703 Rev 05 2018 Page 116 of 134 ...
Страница 117: ...USER MANUAL 16 2000 700 701 702 703 Rev 05 2018 Page 117 of 134 ...
Страница 118: ...USER MANUAL 16 2000 700 701 702 703 Rev 05 2018 Page 118 of 134 ...
Страница 119: ...USER MANUAL 16 2000 700 701 702 703 Rev 05 2018 Page 119 of 134 ...
Страница 120: ...USER MANUAL 16 2000 700 701 702 703 Rev 05 2018 Page 120 of 134 ...
Страница 121: ...USER MANUAL 16 2000 700 701 702 703 Rev 05 2018 Page 121 of 134 ...
Страница 122: ...USER MANUAL 16 2000 700 701 702 703 Rev 05 2018 Page 122 of 134 ...
Страница 123: ...USER MANUAL 16 2000 700 701 702 703 Rev 05 2018 Page 123 of 134 ...
Страница 124: ...USER MANUAL 16 2000 700 701 702 703 Rev 05 2018 Page 124 of 134 ...
Страница 125: ...USER MANUAL 16 2000 700 701 702 703 Rev 05 2018 Page 125 of 134 ...
Страница 126: ...USER MANUAL 16 2000 700 701 702 703 Rev 05 2018 Page 126 of 134 ...
Страница 127: ...USER MANUAL 16 2000 700 701 702 703 Rev 05 2018 Page 127 of 134 ...
Страница 128: ...USER MANUAL 16 2000 700 701 702 703 Rev 05 2018 Page 128 of 134 ...