Operation
45
Offline
Driver not installed
Install driver
Device not listed in the Device manager
of the operating system
Check the USB plug
Device not switched on
Device switched on
Axis motions very noisy
Axis not greased
Grease the axis
Foreign body in the path of the axis
Clean the axis
System is not aligned horizontally
Adjust the system
Door does not fully close
Foreign body in the door gap
Remove the foreign body, press the
Door
close
button twice
Door is not aligned parallel to the hous-
ing
Align the door
Safety circuit open
Door not fully closed, there may be a flat
foreign body in the door gap
Check and clear the door gap, press the
Door close
button
External start is inactive
Incorrect settings in the software
Check and correct the software settings
Problem/Fault
Possible cause
Elimination
NOTICE
Please contact our Technical Customer Service in the first instance if the fault cannot
be eliminated as described above.