Operation
48
Laser will not start
USB cable not connected
properly
Check whether the cable is plugged in correctly.
Interlock circuits closed
Open the interlock circuits (must always be open upon
system start-up).
SD error
Start the marking software and check status / error
reports in the status bar, respectively in the message
window, to receive detailed information.
No laser beam
Shutter closed
Check whether all safety switches are locked. Check
whether the shutter is opened in your marking job.
Defocusing
Check the focus position by changing the distance
between focusing objective and work piece.
Incorrect laser parameters
Check your laser parameters, they may not be suitable
for the work piece to be marked.
Laser power too low
Defocusing
Check the focus position by changing the distance
between focusing objective and work piece.
Incorrect laser parameters
Check your laser parameters, they may not be suitable
for the work piece to be marked.
Laser switches off
automatically
Environmental conditions do
not meet the specifications
Check the environmental conditions.
Keep to the environmental conditions required.
Problem/Fault
Possible cause
Elimination
NOTICE
Please contact our Technical Customer Service in the first instance if the fault
cannot be eliminated as described above.