Operation
56
An overview of the further possible messages and their meaning can be found in the soft-
ware manual.
Interlock fault
Laser Safety Device
SD interlock ERROR
D Cut Off
SD undervoltage
Laser Safety Device
SD ERROR internal under-
voltage
D Cut Off
SD error
Laser Safety Device
SD internal ERROR
D Cut Off
Error +12V, <10V
Laser Interface F
W12V
Display Warning
Error +12V, <9V
Laser Interface F
ERROR +12V
D Cut Off
Error -12V, >-10V
Laser Interface F
WARNING-12V
Display Warning
Error -12V, >-9V
Laser Interface F
ERROR-12V
D Cut Off
Error 24V, <22.5V
Laser Interface F
WARNING 24V
Display Warning
Error 24V, <22.0V
Laser Interface F
ERROR 24V
D Cut Off
Overload I/O
Laser Interface F
ERROR Overload I/O
D Cut Off
Laser over temperature
Laser Interface F
ERROR Over temp. Laser
D Cut Off
BDO not plugged
Scan Head F
ERROR BDO
D Cut Off
Laser head over temperature
Scan Head F
ERROR Overtemp. Shutter
D Cut Off
Galvo error Z
Scan Head F
ERROR Ready Z
Display
Status/Error
Trigger
Message
Evaluation per software