Due to changes in the product, the procedures stated 

can change. This can take place without prior notice 

or adaptation of the operating manual and is no cause 

for complaint, return or price reduction. The user is 

responsible for recognizing discrepancies between 

the operating manual and the actual handling of the 

breathalyzer and must react accordingly.

 

Battery Law notice (BattG):

The device associated with this operating manual 

contains two AAA Alkaline batteries (1.5 V) in delivery 

condition. These batteries are not made for recharging. 

The use of rechargeable batteries is not possible with 

this device and can damage it. If the included

batteries are depleted, they must not be thrown into 

the household  waste. Used batteries may contain 

harmful substances that can cause environmental or 

health damage. Please return all depleted batteries to 

electrical appliance stores, who collect them for free.

Battery exchange:

To replace depleted batteries against new ones, first 

push the battery cover up and then sideways. Then 

replace the depleted batteries with two new ones. Be 

aware of the polarity when inserting the new batteries 

(+/- pole). Put the battery compartment cover on the 

rear panel and push it gently down. 

 

Calibration:

To maintain the accuracy of the breathalyzer, 

ACE recommends a calibration every 6 - 12 months. 

For a calibration service you can send the device 

to ACE Instruments (see www.ace-instruments.net). 

The calibration is not mandatory, but the accuracy 

of the breathalyzer will decrease steadily without it.

We recommend sending it as a parcel, 

so that in cases of loss an inquiry or proof 

of delivery is possible.

FLO

The breath sample was interrupted or not strong 

enough for a measurement. After the sensor 

warmed up itself automatically, please blow 

steadily and without a break into the mouthpiece 

(like you would whistle).

BAT

The battery power is low. Please replace the 

2 batteries inside the device with 2 new ones  

(type AAA, 1.5V, LR03, Alkaline batteries). Be 

aware of the polarity when inserting the new 

batteries (+/- pole). Do not use rechargeable 

batteries!

CAL

If you tested over 1.000 times, the ACE X will 

display “CAL” sign for 3 seconds when warming 

up for test. This “CAL” sign informs about the 

appropriate time for calibration. For a calibration 

service you can send the device to ACE Instru-

ments (see www.ace-instruments.net).

OUT

The device turns off (either by pressing the 

power button constantely for 4 seconds or 10 

seconds after displaying the test result).

C

 or 

A

“C” and “A” is shown when putting on the device 

and keeping pushed the power button.  

C = The number of all tests conducted with the 

device (starting with zero after a calibration)

A = The number of tests where alcohol was 

detected

P

 or 

F

Only displayed in the passive mode, see descrip-

tion in chapter “Conducting the measurement”.